Transcatheter Patent Ductus Arteriosus Closure in Premature Infants: Comparison of Echocardiogram and Angiogram Measurements

被引:1
|
作者
Nijres, Bassel Mohammad [1 ]
Khallaf, Mohamed [1 ]
Bischoff, Adrianne Rahde [2 ]
Carr, Kaitlin [1 ]
Gupta, Umang [1 ]
Mcnamara, Patrick J. [2 ]
Windsor, Jimmy [3 ]
Aldoss, Osamah [1 ]
机构
[1] Univ Iowa, Stead Family Childrens Hosp, Div Pediat Cardiol, BT 1021,200 Hawkins Dr, Iowa City, IA 52242 USA
[2] Univ Iowa, Stead Family Childrens Hosp, Div Neonatol, Iowa City, IA USA
[3] Univ Iowa, Div Pediat Anesthesia, Stead Family Childrens Hosp, Iowa City, IA USA
关键词
Premature infants; PDA closure; PDA dimensions; CATHETER CLOSURE; ACCESS;
D O I
10.1007/s00246-024-03620-8
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Transcatheter patent ductus arteriosus (PDA) closure (TCPC) utilizing transthoracic echocardiogram (TTE) as the sole imaging guide could simplify care. This single-center study compares PDA dimensions obtained from the TTE and angiogram images of patients who underwent attempted TCPC at Stead Family Children's Hospital from 10/01/2019 to 10/31/2020. Blinded investigators measured these dimensions solely for this study and had no impact on clinical care. Also, a hypothetical Piccolo device size was chosen based on the TTE dimensions and another on the angiographic dimensions, and then the correlation was analyzed. Sixty-two patients underwent TCPC attempts. TTE tends to overestimate the PDA narrowest dimension and underestimate the PDA length and aortic end dimension. Linear regression analysis revealed a weak correlation between the length and aortic diameter (R = 0.37 and 0.21, respectively). A modest correlation was observed for the smallest dimension without color Doppler (R = 0.57) and with color Doppler, which was utilized when needed (R = 0.6). Bland-Altman analysis revealed a smaller mean difference between the TTE and angiogram measurements of the narrowest diameter without color Doppler (0.4 mm) and with color Doppler (used as needed) (0.4 mm). However, the mean difference is larger for the aortic end (- 1.64 mm) and the length (- 1.73 mm). TTE accurately predicted the Piccolo device size in 43 (72%) patients and overestimated the size in 17 (28%) patients to the next size. Our findings should be verified with further studies, and additional development of protocols is needed to use TTE to guide TCPC without fluoroscopy.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Can Echocardiogram Predict Ductus Arteriosus Angiogram Measurements in Premature Infants?
    Nijres, Bassel Mohammad
    Khallaf, Mohammad
    Bischoff, Adrianne Rahde
    Carr, Kaitlin
    Gupta, Umang
    McNamara, Patrick McNamara
    Aldoss, Osamah
    PEDIATRIC CARDIOLOGY, 2022, 43 (08) : 1980 - 1981
  • [2] Transcatheter patent ductus arteriosus closure in extremely premature infants
    Garg, Ruchira
    Zahn, Evan
    Sathanandam, Shyam
    Johnson, Jason N.
    PROGRESS IN PEDIATRIC CARDIOLOGY, 2021, 61
  • [3] Transport and Anaesthesia Consideration for Transcatheter Patent Ductus Arteriosus Closure in Premature Infants
    Aw, Tuan Chen
    Chan, Belinda
    Singh, Yogen
    JOURNAL OF CARDIOVASCULAR DEVELOPMENT AND DISEASE, 2023, 10 (09)
  • [4] Transcatheter Patent Ductus Arteriosus Closure in Premature Infants on High Frequency Ventilation
    Dalby, Stephen
    Courtney, Sherry
    Sharma, Megha
    Angtuaco, Michael
    PEDIATRIC CARDIOLOGY, 2022, 43 (08) : 1945 - 1945
  • [5] Transcatheter Closure of Patent Ductus Arteriosus in Under 6 kg and Premature Infants
    Baspinar, Osman
    Sahin, Derya Aydin
    Sulu, Ayse
    Irdem, Ahmet
    Gokaslan, Gokhan
    Sivasli, Ercan
    Kilinc, Metin
    JOURNAL OF INTERVENTIONAL CARDIOLOGY, 2015, 28 (02) : 180 - 189
  • [6] Transcatheter Closure of Patent Ductus Arteriosus in Premature Infants - How Young is Too Young?
    Chakraborty, Abhishek
    Philip, Ranjit
    Naik, Ronak
    Abdulmajeed, Huma
    Harsono, Mimily
    Waller, Rush
    Sathanandam, Shyam
    CIRCULATION, 2020, 142
  • [7] Transcatheter Closure of Patent Ductus Arteriosus in Premature Infants With Very Low Birth Weight
    Wang, Jieh-Neng
    Lin, Yung-Chieh
    Hsieh, Min-Ling
    Wei, Yu-Jen
    Ju, Ying-Tzu
    Wu, Jing-Ming
    FRONTIERS IN PEDIATRICS, 2021, 8
  • [8] Overview of transcatheter patent ductus arteriosus closure in preterm infants
    Almeida-Jones, Myriam
    Tang, Nai Yu
    Reddy, Aneela
    Zahn, Evan
    CONGENITAL HEART DISEASE, 2019, 14 (01) : 60 - 64
  • [9] Transcatheter patent ductus arteriosus closure in premature infants requiring high-frequency ventilation
    Dalby, Stephen T.
    Courtney, Sherry E.
    Sharma, Megha
    Angtuaco, Michael J.
    CARDIOLOGY IN THE YOUNG, 2024, 34 (05) : 1166 - 1168
  • [10] TRANSCATHETER CLOSURE OF PATENT DUCTUS ARTERIOSUS
    高伟
    周爱卿
    余志庆
    李奋
    王荣发
    黄美蓉
    Journal of Shanghai Second Medical University, 2002, (01) : 47 - 49