Modification of Oximeter Ratio to Reduce Pigmentation Bias in Pulse Oximetry

被引:0
|
作者
Arefin, Mohammed Shahriar [1 ]
Shmuylovich, Leonid [2 ,3 ]
Patil, Chetan A. [1 ]
机构
[1] Temple Univ, Dept Bioengn, Philadelphia, PA 19122 USA
[2] Washington Univ, Dept Dermatol, St Louis, MO 14263 USA
[3] Washington Univ, Dept Radiol, St Louis, MO 14263 USA
来源
DESIGN AND QUALITY FOR BIOMEDICAL TECHNOLOGIES XVII | 2024年 / 12833卷
关键词
SKIN COLOR;
D O I
10.1117/12.3003274
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Pulse oximeters are widely used in healthcare systems to estimate blood oxygen saturation level (SpO2) using red and infrared light. Recent clinical and simulation studies reported that in darkly pigmented subjects oximeter over-estimates SpO2 which could lead to higher rates of occult hypoxemia in highly pigmented subjects. The probable solutions to solve this over-estimation bias could be modification of the current oximeter design, calibration enrollment or modification of oximeter ratio (R). In this study, a modification of the current oximeter ratio (R) was presented by using different combinations of currently estimated oximeter parameters. Simulation results showed that modified oximeter ratio reduces over-estimation bias in highly pigmented subjects compared to the conventional oximeter ratio. In the regions near hypoxemia threshold (90% oxygenation level), the over-estimation bias in the simulated test cohort could reduce from 1.36% to -0.01% if modified oximeter ratio is used. Results show that modification of oximeter ratio could be used in future to improve oximeter accuracy and produce pigmentation independent outcomes.
引用
收藏
页数:5
相关论文
共 50 条
  • [1] Assessment of skin pigmentation-related bias in pulse oximetry readings among adults
    Ashish K. Khanna
    John Beard
    Sakari Lamminmäki
    Johanna Närväinen
    Nicholas Antaki
    Halit O. Yapici
    Journal of Clinical Monitoring and Computing, 2024, 38 : 113 - 120
  • [2] Assessment of skin pigmentation-related bias in pulse oximetry readings among adults
    Khanna, Ashish K.
    Beard, John
    Lamminmaki, Sakari
    Narvainen, Johanna
    Antaki, Nicholas
    Yapici, Halit O.
    JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2024, 38 (01) : 113 - 120
  • [3] Scaling of Algorithmic Bias in Pulse Oximetry with Signal-to-Noise Ratio
    Cao, Jiaming
    Mehta, Neil Ashim
    Wu, Jingyi
    Wood, Sossena
    Kainerstorfer, Jana M.
    Grover, Pulkit
    2023 45TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY, EMBC, 2023,
  • [4] Pulse oximetry, racial bias and statistical bias: further improvements of pulse oximetry are necessary
    Marco Ferrari
    Valentina Quaresima
    Felix Scholkmann
    Annals of Intensive Care, 12
  • [5] Pulse oximetry, racial bias and statistical bias: further improvements of pulse oximetry are necessary
    Ferrari, Marco
    Quaresima, Valentina
    Scholkmann, Felix
    ANNALS OF INTENSIVE CARE, 2022, 12 (01)
  • [6] POTENTIAL ERRORS IN PULSE OXIMETRY .1. PULSE OXIMETER EVALUATION
    RALSTON, AC
    WEBB, RK
    RUNCIMAN, WB
    ANAESTHESIA, 1991, 46 (03) : 202 - 206
  • [7] A review of the effect of skin pigmentation on pulse oximeter accuracy
    Al-Halawani, Raghda
    Charlton, Peter H.
    Qassem, Meha
    Kyriacou, Panayiotis A.
    PHYSIOLOGICAL MEASUREMENT, 2023, 44 (05)
  • [8] Pulse oximetry, racial bias and statistical bias
    Martin J. Tobin
    Amal Jubran
    Annals of Intensive Care, 12
  • [9] Pulse oximetry, racial bias and statistical bias
    Tobin, Martin J.
    Jubran, Amal
    ANNALS OF INTENSIVE CARE, 2022, 12 (01)
  • [10] A REVIEW OF THE PRINCIPLES OF PULSE OXIMETRY AND ACCURACY OF PULSE OXIMETER ESTIMATES DURING EXERCISE
    MENGELKOCH, LJ
    MARTIN, D
    LAWLER, J
    PHYSICAL THERAPY, 1994, 74 (01): : 40 - 49