Novel multileaf collimator designs with tongues and grooves in leaf end and Monte Carlo simulations

被引:0
|
作者
Cui, Weijie [1 ]
Tian, Yuan [1 ]
Dai, Jianrong [1 ]
机构
[1] Chinese Acad Med Sci & Peking Union Med Coll, Canc Hosp, Natl Canc Ctr, Dept Radiat Oncol, Beijing 100021, Peoples R China
关键词
Multileaf collimator; Leaf end; Leaf gap; Tongue and groove; SYSTEM; IMRT; GAP;
D O I
10.1016/j.meddos.2024.01.008
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
In this study, we proposed 2 new multileaf collimator leaf designs to eliminate leaf gaps for closed leaf pairs so that radiation leakage can be avoided. In the new designs, multi tongues and grooves were added to the conventional multileaf collimators leaf ends. Thus, when a pair of leaves closed, tongues of a leaf can enter grooves of its opposing leaf. Consequently, there would be no radiation leakage through closed leaves. One design was named finger-shaped MLC, and another design with doubled leaf end thickness was named hand-shaped MLC. Monte Carlo simulations were performed to simulate dosimetric characteristics of the new MLC designs and comparison to conventional MLCs was performed. The simulations show that for the closed field, the new designs reduce leakage dramatically. And for the open field, the finger-shaped MLC has a larger penumbra width than conventional MLC, while the penumbra for the hand-shaped MLC is comparable to that of conventional MLC. With the application of new MLC designs, it is expected to eliminate leaf gaps for MLC usage and protect normal tissues better. (c) 2024 Published by Elsevier Inc. on behalf of American Association of Medical Dosimetrists.
引用
收藏
页码:254 / 262
页数:9
相关论文
共 50 条
  • [31] Characterizing a Geant4 Monte Carlo model of a multileaf collimator for a TrueBeam™ linear accelerator
    Kandlakunta, Praneeth
    Momin, Shadab
    Sloop, Austin
    Zhang, Tiezhi
    Khan, Rao
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 59 : 1 - 12
  • [32] Lateral Penumbra Modelling Based Leaf End Shape Optimization for Multileaf Collimator in Radiotherapy
    Zhou, Dong
    Zhang, Hui
    Ye, Peiqing
    COMPUTATIONAL AND MATHEMATICAL METHODS IN MEDICINE, 2016, 2016
  • [33] A Novel Dynamic Multileaf Collimator Leaf-Sequencing Algorithm in Radiation Therapy
    Jing, J.
    Lin, H.
    Chow, J.
    MEDICAL PHYSICS, 2016, 43 (06) : 3584 - 3584
  • [34] Monte Carlo simulation of multileaf-collimators with curved leaf-ends
    Haryanto, F.
    Fippel, M.
    Laub, W.
    Nüsslin, F.
    Zeitschrift fur Medizinische Physik, 2001, 11 (03): : 172 - 178
  • [35] Modeling a Multi Leaf Collimator for IMRT Monte Carlo Dose Calculations
    Craig, J.
    Mulligan, M.
    Chen, J.
    Wong, E.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [36] Optimization PEM designs based on Monte Carlo simulations
    Wang, Mengdie
    Liu, Yilong
    Hu, Guangshu
    Zhang, Hui
    Qinghua Daxue Xuebao/Journal of Tsinghua University, 2014, 54 (08): : 1092 - 1097
  • [37] X-RAY COLLIMATOR DESIGN USING MONTE CARLO SIMULATIONS
    Han, Eun Young
    Clarkson, Richard
    Sharma, Sunil
    Corry, Peter M.
    Moros, Eduardo G.
    Ratanatharathorn, Vaneerat
    BIOMEDICAL ENGINEERING-APPLICATIONS BASIS COMMUNICATIONS, 2013, 25 (06):
  • [38] Monte Carlo simulation of the dynamic micro-multileaf collimator of a LINAC Elekta Precise using PENELOPE
    Gonzalez, W.
    Lallena, A. M.
    Alfonso, R.
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (11): : 3417 - 3431
  • [39] Novel Monte Carlo dose calculation algorithm for robotic radiosurgery with multi leaf collimator: Dosimetric evaluation
    Heidorn, Sarah-Charlotta
    Kilby, Warren
    Fuerweger, Christoph
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 55 : 25 - 32
  • [40] Suggesting a new design for multileaf collimator leaves based on Monte Carlo simulation of two commercial systems
    Hariri, Sanaz
    Shahriari, Majid
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2010, 11 (03): : 173 - 185