共 31 条
A FLEXIBLE HAIR-LIKE LASER INDUCED GRAPHITIC SENSOR FOR LOW FLOW RATE SENSING APPLICATIONS
被引:0
|作者:
Abbasnejad, Behrokh
[1
]
McGloin, David
[1
]
Clemon, Lee
[2
]
机构:
[1] Univ Technol Sydney, Sch Elect & Data Engn, Sydney, NSW, Australia
[2] Univ Technol Sydney, Sch Mech & Mechatron Engn, Sydney, NSW, Australia
来源:
关键词:
FAILURE MODE;
D O I:
暂无
中图分类号:
TP18 [人工智能理论];
学科分类号:
081104 ;
0812 ;
0835 ;
1405 ;
摘要:
Direct low flow sensing is of interest to many applications in medical and biochemical industries. Low flow rate measurement is still challenging, and conventional flow sensors such as hot films, hot wires and Pitot probes are not capable of measuring very low flow rates accurately. In some applications that require flow measurement in a small diameter tubing (e.g. intravenous (IV) infusion), using such sensors also becomes mechanically impractical. Herein, a flexible laser-induced graphitic (LIG) piezoresistive flow sensor has been fabricated in a cost-effective single processing step. The capability of the LIG sensor in very low flow rate measurement has been investigated by embedding the sensor within an intravenous (IV) line. The embedded LIG hair-like sensor was tested at ambient temperature within the IV line at flow rates ranging from 0 m/s to 0.3 m/s (IV infusion free-flow rate). The LIG hair-like sensor presented in this study detects live flow rates of IV infusions with a threshold detection limit as low as 0.02 m/s. Moreover, the deformation of the LIG hair-like sensor that lead to resistance change in response to various flow rates is simulated using COMSOL Multiphysics.
引用
收藏
页数:5
相关论文