An assessment of the ageing of the absolute spectral responsivity of trap detectors at krypton ion laser emission wavelengths

被引:0
|
作者
Dury, Martin [1 ]
Harris, Subrena [1 ]
Goodman, Teresa [1 ]
Servantes, William [1 ]
White, Malcolm [2 ,3 ]
机构
[1] Natl Phys Lab, Opt Radiometr Metrol Grp, Teddington, England
[2] Univ Colorado, Dept Phys, Boulder, CO 80309 USA
[3] Natl Inst Stand & Technol, Boulder, CO 80305 USA
关键词
radiometry; responsivity; cryogenic radiometer; trap detector;
D O I
10.1088/1681-7575/ad615d
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We have assessed the absolute spectral responsivities of three-element trap detectors calibrated against the National Physical Laboratory primary standard cryogenic radiometer at eight krypton ion laser emission wavelengths over an approximate period of 20 years. As expected, we observed that the absolute spectral responsivities decrease with time. Additionally, we found that as the wavelength decreases, the ageing rate increases. At 356.4 nm, an average responsivity decrease of around -1800 ppm yr-1 is reported. While at longer wavelengths, the slowest rate of ageing, approximately -40 ppm yr-1, occurs at 799.3 nm. Interestingly, the year-on-year changes exhibit a relatively linear trend, allowing us to predict the future performance of the trap detectors. It is important to note, however, that predicted performance cannot be guaranteed and regular recalibration against a primary standard is essential, especially if low calibration uncertainties are required.
引用
收藏
页数:9
相关论文
共 42 条
  • [1] Influence of vignetting errors on the relative spectral responsivity of trap detectors
    Stock, KD
    Heine, R
    METROLOGIA, 1998, 35 (04) : 447 - 450
  • [2] Influence of vignetting errors on the relative spectral responsivity of trap detectors
    Phys.-Technische Bundesanstalt, Bundesallee 100, D-38116 Braunschweig, Germany
    Metrologia, 4 (447-450):
  • [3] Absolute spectral responsivity of silicon trap detector based on absolute cryogenic radiometer
    Shi, Xueshun
    Liu, Changming
    Liu, Yulong
    Yang, Lechen
    Zhao, Kun
    Chen, Haidong
    INTERNATIONAL CONFERENCE ON PHOTONICS AND OPTICAL ENGINEERING (ICPOE 2014), 2015, 9449
  • [4] Ultraviolet spectral responsivity of silicon trap detectors traceable to a cryogenic radiometer
    Liu C.-M.
    Shi X.-S.
    Chen H.-D.
    Liu Y.-L.
    Zhao K.
    Ying C.-P.
    Chen K.-F.
    Li L.-G.
    Shi, Xue-Shun (xshshi@163.com), 2016, Chinese Optical Society (45):
  • [5] Characterization of silicon photodiode-based trap detectors and establishment of spectral responsivity scale
    Bazkir, Ö
    Samadov, F
    OPTICS AND LASERS IN ENGINEERING, 2005, 43 (02) : 131 - 141
  • [6] Accurate calibration of the spectral responsivity of silicon trap detectors between 350 nm and 1064 nm
    Li Jian-Jun
    Zheng Xiao-Bing
    Lu Yun-Jun
    Zhang-Wei
    Xie Ping
    Zou Peng
    ACTA PHYSICA SINICA, 2009, 58 (09) : 6273 - 6278
  • [7] Absolute irradiance responsivity calibration using diode lasers emitting at three wavelengths for tricolor laser applications
    Tanabe, Minoru
    Kinoshita, Kenichi
    OPTIK, 2020, 202
  • [8] Picosecond laser krypton plasma emission in water window spectral range
    Vrba, P.
    Vrbova, M.
    Mueller, M.
    Mann, K.
    Panek, D.
    Parkman, T.
    PHYSICS OF PLASMAS, 2017, 24 (12)
  • [9] High accuracy measurement of the absolute spectral responsivity of Ge and InGaAs trap detectors by direct calibration against an electrically calibrated cryogenic radiometer in the near-infrared
    Lopez, M.
    Hofer, H.
    Kueck, S.
    METROLOGIA, 2006, 43 (06) : 508 - 514
  • [10] Accurate determination of the spectral responsivity of silicon trap detectors between 238 nm and 1015 nm using a laser-based cryogenic radiometer
    Werner, L
    Fischer, J
    Johannsen, U
    Hartmann, J
    METROLOGIA, 2000, 37 (04) : 279 - 284