An artificial neural network based approach for predicting the proton beam spot dosimetric characteristics of a pencil beam scanning technique

被引:1
|
作者
Ranjith, C. P. [1 ,2 ]
Krishnan, Mayakannan [1 ]
Raveendran, Vysakh [2 ]
Chaudhari, Lalit [2 ]
Laskar, Siddhartha [2 ]
机构
[1] DY Patil Educ Soc Deemed Be Univ, Ghaziabad, India
[2] Homi Bhabha Natl Inst, Adv Ctr Treatment Res & Educ Canc, Dept Radiat Oncol, Mumbai, Maharashtra, India
关键词
machine learning; pencil beam scanning; quality assurance; log file; proton therapy; HEAD;
D O I
10.1088/2057-1976/ad3ce0
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Utilising Machine Learning (ML) models to predict dosimetric parameters in pencil beam scanning proton therapy presents a promising and practical approach. The study developed Artificial Neural Network (ANN) models to predict proton beam spot size and relative positional errors using 9000 proton spot data. The irradiation log files as input variables and corresponding scintillation detector measurements as the label values. The ANN models were developed to predict six variables: spot size in the x-axis, y-axis, major axis, minor axis, and relative positional errors in the x-axis and y-axis. All ANN models used a Multi-layer perception (MLP) network using one input layer, three hidden layers, and one output layer. Model performance was validated using various statistical tools. The log file recorded spot size and relative positional errors, which were compared with scintillator-measured data. The Root Mean Squared Error (RMSE) values for the x-spot and y-spot sizes were 0.356 mm and 0.362 mm, respectively. Additionally, the maximum variation for the x-spot relative positional error was 0.910 mm, while for the y-spot, it was 1.610 mm. The ANN models exhibit lower prediction errors. Specifically, the RMSE values for spot size prediction in the x, y, major, and minor axes are 0.053 mm, 0.049 mm, 0.053 mm, and 0.052 mm, respectively. Additionally, the relative spot positional error prediction model for the x and y axes yielded maximum errors of 0.160 mm and 0.170 mm, respectively. The normality of models was validated using the residual histogram and Q-Q plot. The data over fit, and bias were tested using K (k = 5) fold cross-validation, and the maximum RMSE value of the K fold cross-validation among all the six ML models was less than 0.150 mm (R-Square 0.960). All the models showed excellent prediction accuracy. Accurately predicting beam spot size and positional errors enhances efficiency in routine dosimetric checks.
引用
收藏
页数:13
相关论文
共 50 条
  • [1] Dosimetric Comparison of Various Spot Placement Techniques in Proton Pencil Beam Scanning
    Rehman, Mahboob ur
    Zeidan, Omar A.
    Willoughby, Twyla
    Meeks, Sanford L.
    Kelly, Patrick
    Erhart, Kevin
    INTERNATIONAL JOURNAL OF PARTICLE THERAPY, 2022, 9 (01) : 54 - 63
  • [2] An optimized approach for robust spot placement in proton pencil beam scanning
    Rehman, Mahboob Ur
    Erhart, Kevin
    Kielbasa, Jerrold
    Meeks, Sanford L.
    Li, Zhiqiu
    Willoughby, Twyla
    Ramakrishna, Naren
    Stephenson, Ken
    Rahman, Talat S.
    Kelly, Patrick
    Zeidan, Omar
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (23):
  • [3] Spot Position Uncertainty of Scanning Proton Pencil Beam
    Li, H.
    Zhu, X.
    Ciangaru, G.
    Sahoo, N.
    Suzuki, K.
    Lee, A.
    Gillin, M.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [4] An Optimized Approach for Spot Placement in Pencil Beam Scanning (PBS) Proton Therapy
    Rehman, M.
    Erhart, K.
    Meeks, S.
    Kelly, P.
    Zeidan, O.
    MEDICAL PHYSICS, 2019, 46 (06) : E140 - E141
  • [5] Deriving Spot Shape Criteria for Proton Pencil Beam Scanning
    Wulff, J.
    Huggins, A.
    MEDICAL PHYSICS, 2016, 43 (06) : 3652 - 3652
  • [6] Dosimetric response of radiophotoluminescent glass detectors in a scanning pencil proton beam
    Pasaricek, Luka
    Knezevic, Zeljka
    Brkic, Hrvoje
    Davidkova, Marie
    Navratil, Matej
    Vondracek, Vladimir
    Majer, Marija
    RADIATION MEASUREMENTS, 2024, 176
  • [7] A Stochastic Approach to Daily QA Tolerances On Spot Properties for Proton Pencil Beam Scanning
    James, S. St.
    Saini, J.
    Bloch, C.
    MEDICAL PHYSICS, 2016, 43 (06) : 3503 - 3503
  • [8] A Knowledge-Based Planning Approach for Dosimetric Prediction of Prostate Proton Pencil-Beam Scanning Plans
    Zou, W.
    Teo, B.
    O'Reilly, S.
    Ainsley, C.
    Lin, A.
    Xiao, Y.
    Dong, L.
    Yin, L.
    MEDICAL PHYSICS, 2018, 45 (06) : E658 - E658
  • [9] Factors Predicting Dosimetric Consequence of Gastrointestinal Density in Pencil Beam Scanning Proton Therapy for Abdominal Neuroblastoma
    Lim, P. S.
    Rompokos, V.
    Bizzocchi, N.
    Gillies, C.
    Gosling, A.
    Royle, G.
    Chang, Y. -C.
    Gains, J.
    Gaze, M.
    PEDIATRIC BLOOD & CANCER, 2020, 67 : S73 - S74
  • [10] Pencil Beam Scanning:: a dynamical approach to proton therapy
    Sepulchre, R
    Gérard, M
    Marchand, B
    Prieels, D
    Bauvir, B
    IEEE INTERNATIONAL SYMPOSIUM ON BIO-INFORMATICS AND BIOMEDICAL ENGINEERING, PROCEEDINGS, 2000, : 370 - 375