Blood Pressure Measurement Device Accuracy Evaluation: Statistical Considerations with an Implementation in R

被引:2
|
作者
Chandel, Tanvi [1 ]
Miranda, Victor [2 ]
Lowe, Andrew [1 ]
Lee, Tet Chuan [1 ]
机构
[1] Auckland Univ Technol, Inst Biomed Technol, Auckland 1010, New Zealand
[2] Auckland Univ Technol, Dept Math Sci, Auckland 1010, New Zealand
关键词
blood pressure; ANSI/AAMI-SP10; standards; blood pressure measurement device; probability of acceptance; HYPERTENSION-INTERNATIONAL-PROTOCOL; PULSE TRANSIT-TIME; EUROPEAN-SOCIETY; VALIDATION;
D O I
10.3390/technologies12040044
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Inaccuracies from devices for non-invasive blood pressure measurements have been well reported with clinical consequences. International standards, such as ISO 81060-2 and the seminal AAMI/ANSI SP10, define protocols and acceptance criteria for these devices. Prior to applying these standards, a sample size of N >= 85 is mandatory, that is, the number of distinct subcjects used to calculate device inaccuracies. Often, it is not possible to gather such a large sample. Many studies apply these standards with a smaller sample. The objective of the paper is to introduce a methodology that broadens the method first developed by the AAMI Sphygmomanometer Committee for accepting a blood pressure measurement device. We study changes in the acceptance region for various sample sizes using the sampling distribution for proportions and introduce a methodology for estimating the exact probability of the acceptance of a device. This enables the comparison of the accuracies of existing device development techniques even if they were studied with a smaller sample size. The study is useful in assisting BP measurement device manufacturers. To assist clinicians, we present a newly developed "bpAcc" package in R to evaluate acceptance statistics for various sample sizes.
引用
收藏
页数:17
相关论文
共 50 条
  • [31] Perinatal Blood Pressure Measurement: A Critical Need for Accuracy. Implementation of an Evidence-Based Protocol
    Young, Roselyn
    JOGNN-JOURNAL OF OBSTETRIC GYNECOLOGIC AND NEONATAL NURSING, 2013, 42 : S63 - S63
  • [32] DESIGN AND EVALUATION OF A DEVICE FOR MEASUREMENT OF INTERFACE PRESSURE
    STEINBERG, MD
    COOKE, ED
    JOURNAL OF BIOMEDICAL ENGINEERING, 1993, 15 (06): : 464 - 468
  • [33] Evaluation of a continuous non-invasive arterial blood pressure monitoring device in comparison with an arterial blood pressure measurement in the ICU
    K Smolle
    M Schmid
    Critical Care, 15 (Suppl 1):
  • [34] Low Cost Field Test Measurement Method and Prototype Measurement Device Implementation for Timing Accuracy Evaluation of IEEE 1588 Solutions
    Kovacshazy, Tamas
    Hollos, Adam Erik
    2018 IEEE INTERNATIONAL WORKSHOP ON METROLOGY FOR INDUSTRY 4.0 AND IOT (METROIND4.0&IOT), 2018, : 72 - 77
  • [35] Factors influencing the accuracy of home blood pressure measurement
    Merrick, RD
    Olive, KE
    SOUTHERN MEDICAL JOURNAL, 1997, 90 (11) : 1110 - 1114
  • [36] Accuracy improvement of noninvasive finger blood pressure measurement
    Lackovic, I
    Santic, A
    MEDICON 2001: PROCEEDINGS OF THE INTERNATIONAL FEDERATION FOR MEDICAL & BIOLOGICAL ENGINEERING, PTS 1 AND 2, 2001, : 281 - 284
  • [37] No effect of sleeved arms on the accuracy of blood pressure measurement
    Li, Ya
    Li, Yi
    Li, Fang
    Liu, Demin
    Zhang, Yanan
    Cui, Xiaoran
    Cui, Wei
    INTERNAL MEDICINE JOURNAL, 2021, 51 (12) : 2087 - 2094
  • [38] THE ACCURACY OF AUSCULTATORY BLOOD-PRESSURE MEASUREMENT IN THE ELDERLY
    NIELSEN, PE
    ACTA MEDICA SCANDINAVICA, 1983, : 39 - 44
  • [39] Accuracy of indirect measurement of blood pressure in neonatal foals
    Giguère, S
    Knowles, HA
    Valverde, A
    Bucki, E
    Young, L
    JOURNAL OF VETERINARY INTERNAL MEDICINE, 2005, 19 (04) : 571 - 576
  • [40] Accuracy of oscillometric blood pressure measurement in atrial fibrillation
    Feenstra, Rudolf K.
    Allaart, Cor P.
    Berkelmans, Gijs F. N.
    Westerhof, Berend E.
    Smulders, Yvo M.
    BLOOD PRESSURE MONITORING, 2018, 23 (02) : 59 - 63