Discrepancy between arterial oxygen saturation and pulse oximetry measurement in a Chinese pediatric patient cohort

被引:0
|
作者
Chen, Qin-chang [1 ]
Shen, Jun-jun [1 ]
Huang, Yu-lu [1 ]
Kong, Ran [1 ]
Xie, Yu-mei [1 ]
Wang, Shu- shui [1 ]
机构
[1] Southern Med Univ, Guangdong Prov Peoples Hosp, Guangdong Acad Med Sci, Dept Pediat Cardiol, Guangzhou 510000, Peoples R China
关键词
Oximetry; Oxygen saturation; Racial groups; Occult hypoxemia; Pediatrics; SKIN PIGMENTATION; ACCURACY; RELIABILITY;
D O I
10.1016/j.heliyon.2024.e28336
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Background: Increasing evidence suggest a racial bias in pulse oximetry measurement, but this was under investigated in Asian pediatric populations. Methods: Via the Pediatric Intensive Care database, this retrospective study included pediatric patient records of arterial oxygen saturation (SaO(2)) and oxygen saturation on pulse oximetry (SpO(2)) measured within 10 min. Discrepancy was examined, and potential predictors of occult hypoxemia (defined as SaO(2) <88% with the paired SpO(2) >= 92%) as well as its association with outcomes were explored by logistic regression. Results: A total of 390 patients were included with 454 pairs of SaO(2)-SpO(2) readings. The study population consisted of Han Chinese (99.0%) and 43.6% were female. Occult hypoxemia was observed in 20.0% of the patients, with a mean SaO(2) of 71.4 +/- 15.8%. Potential predictors of occult hypoxemia included female, being first admitted to cardiac ICU, congenital heart disease, increased heart rate, while patients with prior surgery records were less likely to experience occult hypoxemia. Patients with occult hypoxemia had numerically higher in-ICU mortality (16.7% versus 10.9%) and in-hospital mortality (17.9% versus 10.9%), but the associations were not statistically significant. Conclusions: There was a substantial proportion of hypoxemia that was not detected by pulse oximetry in the Chinese pediatric patients, which might be predicted by several characteristics and seemed to associate with mortality.
引用
收藏
页数:10
相关论文
共 50 条
  • [31] Analysis of Discrepancies Between Pulse Oximetry and Arterial Oxygen Saturation Measurements by Race and Ethnicity and Association With Organ Dysfunction and Mortality
    Wong, An-Kwok Ian
    Charpignon, Marie
    Kim, Han
    Josef, Christopher
    de Hond, Anne A. H.
    Fojas, Jhalique Jane
    Tabaie, Azade
    Liu, Xiaoli
    Mireles-Cabodevila, Eduardo
    Carvalho, Leandro
    Kamaleswaran, Rishikesan
    Madushani, R. W. M. A.
    Adhikari, Lasith
    Holder, Andre L.
    Steyerberg, Ewout W.
    Buchman, Timothy G.
    Lough, Mary E.
    Celi, Leo Anthony
    JAMA NETWORK OPEN, 2021, 4 (11)
  • [32] Correlation between hepatopulmonary syndrome and oxygen saturation pulse oximetry in cirrhotic patients
    Paes Rose, Sara Cardoso
    Cunha, Debora Vilela
    Cobra Medeiros, Sandra de Barros
    Trevizoli, Jose Eduardo
    Carneiro, Marcos de Vasconcelos
    de Freitas, Wladimir Magalhaes
    Costa Mendes, Liliana Sampaio
    REVISTA DA ASSOCIACAO MEDICA BRASILEIRA, 2020, 66 (11): : 1577 - 1582
  • [33] Correlation and concordance of arterial oxygen saturation measured by arterial blood gases and pulse oximetry according to anatomical location
    Correa, Cristhian
    Gallo Gonzalez, Virginia
    Moreno-Tome, Rosa Maria Rojas
    Suarez Rubio, Maria Del Carmen
    Pesquera Criado, Rafael
    Lopez Padilla, Daniel
    Antonella Benedetti, Paola
    Garcia Lopez, Jose Javier
    Puente Maestu, Luis
    EUROPEAN RESPIRATORY JOURNAL, 2023, 62
  • [34] CONTINUOUS MONITORING OF ARTERIAL OXYGEN-SATURATION WITH PULSE OXIMETRY DURING TRANSFER TO THE RECOVERY ROOM
    TYLER, IL
    TANTISIRA, B
    WINTER, PM
    MOTOYAMA, EK
    ANESTHESIA AND ANALGESIA, 1985, 64 (11): : 1108 - 1112
  • [35] Effect of skin tone on the accuracy of the estimation of arterial oxygen saturation by pulse oximetry: a systematic review
    Martin, Daniel
    Johns, Chris
    Sorrell, Lexy
    Healy, Eugene
    Phull, Mandeep
    Olusanya, Segun
    Peters, Mark
    Fabes, Jeremy
    BRITISH JOURNAL OF ANAESTHESIA, 2024, 132 (05) : 945 - 956
  • [36] Accuracy of pulse oximetry in preterm and term infants is insufficient to determine arterial oxygen saturation and tension
    Wackernagel, Dirk
    Blennow, Mats
    Hellstrom, Ann
    ACTA PAEDIATRICA, 2020, 109 (11) : 2251 - 2257
  • [37] INTRAPARTUM FETAL PULSE OXIMETRY - THE EFFECTS OF MATERNAL HYPEROXIA ON FETAL ARTERIAL OXYGEN-SATURATION
    DILDY, GA
    CLARK, SL
    LOUCKS, CA
    AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1994, 171 (04) : 1120 - 1124
  • [38] CLINICAL-RESULTS OF FETAL PULSE OXIMETRY MEASURING THE FETAL ARTERIAL OXYGEN-SATURATION
    BUTTERWEGGE, M
    LAMPE, S
    GEBURTSHILFE UND FRAUENHEILKUNDE, 1995, 55 (11) : 634 - 641
  • [39] Extracorporeal Membrane Oxygenation Physiological Factors Influence Pulse Oximetry and Arterial Oxygen Saturation Discrepancies
    Kalra, Andrew
    Shou, Benjamin L.
    Zhao, David
    Wilcox, Christopher
    Keller, Steven P.
    Kim, Bo Soo
    Whitman, Glenn J. R.
    Cho, Sung-Min
    ANNALS OF THORACIC SURGERY, 2024, 117 (06): : 1221 - 1228
  • [40] NASAL PULSE OXIMETRY OVERESTIMATES OXYGEN-SATURATION
    ROSENBERG, J
    PEDERSEN, MH
    ANAESTHESIA, 1990, 45 (12) : 1070 - 1071