THE PERFORMANCE OF MINI WRIGHT PEAK FLOW METERS AFTER PROLONGED USE

被引:12
|
作者
MILES, JF
BRIGHT, P
AYRES, JG
CAYTON, RM
MILLER, MR
机构
[1] UNIV BIRMINGHAM,GOOD HOPE NHS TRUST,DEPT MED,BIRMINGHAM B75 7RR,W MIDLANDS,ENGLAND
[2] BIRMINGHAM HEARTLANDS HOSP,CHEST RES INST,BIRMINGHAM B9 5ST,W MIDLANDS,ENGLAND
关键词
D O I
10.1016/0954-6111(95)90227-9
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The accuracy of 84 new and 35 old Mini Wright peak flow meters were tested using a servo-controlled pump system. The 95% confidence limits for flow measurement across the range of the new meters was between +/- 151 min(-1) at the lower end of the range and +/-281 min(-1) at the top of the range. The readings for 22 (63%) of the old meters (age range 1-13 yr) were within these 95% confidence limits. For the remaining 13 old meters (age range 1-13 yr) whose readings were not within these limits, there were 11 meters with readings falling below and two meters with readings above these limits. Twelve of these old meters were washed and retested and there was no significant change in their readings. Twenty of the new meters were retested after 1 yr of continuous use and their readings were significantly higher with a median value of 51 min(-1) across the range, although only two of these 20 meters had readings outside the 95% confidence limits set from the 84 new meters. It is concluded that whilst Mini Wright meters aged up to 14 yr can give readings which are as good as new meters, some meters demonstrate significant changes in readings after only 1 yr and washing did not correct this change. It is recommended that clinicians prescribing peak expiratory flow (PEF) meters should be responsible for checking the patient's meter as well as their PEF readings at clinic visits.
引用
收藏
页码:603 / 605
页数:3
相关论文
共 50 条
  • [31] Gender differences in the use of peak flow meters and their effect on peak expiratory flow
    Self, TH
    Cross, LB
    Nolan, SE
    Weibel, JB
    Hilaire, M
    Franks, AR
    Finch, CK
    Tolley, EA
    PHARMACOTHERAPY, 2005, 25 (04): : 526 - 530
  • [32] BETTER MANAGEMENT OF ASTHMA WITH MINI-WRIGHT PEAK FLOW METER
    KATZ, DN
    WESTERN JOURNAL OF MEDICINE, 1982, 137 (01): : 54 - 55
  • [33] ASSESSMENT OF THE PERFORMANCE OF 2 MINI-FLOW METERS
    TECULESCU, DB
    HANNHART, B
    MALVESTIO, P
    BAHLAOUI, A
    SADOUL, P
    REVUE DES MALADIES RESPIRATOIRES, 1987, 4 (03) : 121 - 125
  • [34] PEAK FLOW METERS
    KAWANE, H
    SOEJIMA, R
    KOJIMA, K
    BRITISH JOURNAL OF GENERAL PRACTICE, 1992, 42 (357): : 168 - 168
  • [35] Peak flow meters
    Komesaroff, D
    NEW ZEALAND MEDICAL JOURNAL, 1997, 110 (1038) : 63 - 63
  • [36] PEAK FLOW METERS
    FAHY, BF
    BACHUS, BF
    CHEST, 1983, 83 (04) : 710 - 710
  • [37] Peak Expiratory Flow in Normals: Comparison of the Mini Wright versus Spirometric Predicted Peak Flows
    Pesola, Gene R.
    O'Donnell, Pamela
    Pesola, Gene R., Jr.
    Chinchilli, Vernon M.
    Saari, Arthur F.
    JOURNAL OF ASTHMA, 2009, 46 (08) : 845 - 848
  • [38] Performance testing for different peak expiratory flow meters
    Wu, Zhongping
    Huang, Ruibo
    Zhong, Liping
    Zheng, Jinping
    Gao, Yi
    TECHNOLOGY AND HEALTH CARE, 2023, 31 (01) : 141 - 149
  • [39] MEASURING PEAK EXPIRATORY FLOW IN GENERAL-PRACTICE - COMPARISON OF MINI WRIGHT PEAK FLOW METER AND TURBINE SPIROMETER
    JONES, KP
    MULLEE, MA
    BRITISH MEDICAL JOURNAL, 1990, 300 (6740): : 1629 - 1631
  • [40] Use of peak expiratory flow meters in adults: Practical aspects
    Sears, MR
    EUROPEAN RESPIRATORY JOURNAL, 1997, 10 : S72 - S74