DEVICE INTERACTION - ANTITACHYCARDIA PACEMAKERS AND DEFIBRILLATORS FOR SUSTAINED VENTRICULAR-TACHYCARDIA

被引:8
|
作者
AHERN, TS
NYDEGGER, C
MCCORMICK, DJ
MARINCHAK, R
KOWEY, P
HOROWITZ, LN
WORLEY, S
KUTALEK, SP
机构
[1] PHILADELPHIA ARRHYTHMIA GRP,PHILADELPHIA,PA
[2] HAHNEMANN UNIV,PHILADELPHIA,PA 19102
[3] LANKENAU HOSP,PHILADELPHIA,PA 19151
[4] PRESBYTERIAN UNIV PENN MED CTR,PHILADELPHIA,PA
[5] LANCASTER GEN HOSP,LANCASTER,PA
来源
关键词
ANTITACHYCARDIA PACEMAKERS; DEFIBRILLATORS; DEVICE INTERACTION; SUSTAINED VENTRICULAR TACHYCARDIA;
D O I
10.1111/j.1540-8159.1991.tb05111.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We evaluated the combined use of permanent automatic antitachycardia pacemakers and implanted defibrillators in ten patients with recurrent monomorphic sustained ventricular tachycardia (VT). Pacemaker programming was VVI-T automatic burst in eight patients, VVI-T magnet mode in one patient, and VVI in one patient. Device interactions occurred in four patients, requiring changes in pacemaker programming. These included defibrillator multiple counting during pacing, inappropriate pacemaker bursts initiating VT, inappropriate reset of the pacemaker antitachycardia mode by defibrillation, defibrillator discharge after pacemaker VT termination, and defibrillator VT reinitiation. Two patients required pacemaker programming out of the antitachycardia mode, and two required a change in antitachycardia pacing parameters. Seven patients remain in automatic VVI-T and three in VVI modes. Mean follow-up is 13 months and all patients are alive. Thus, although pacemaker/defibrillator combinations function well for patients with more than one VT rate, device interactions occur frequently and may require pacemaker reprogramming or elimination of the overdrive mode. Combined use of these devices should be cautiously considered when single device therapy is unsatisfactory. Devices that combine both pacing and defibrillation features may reduce adverse interaction.
引用
收藏
页码:302 / 307
页数:6
相关论文
共 50 条
  • [21] INITIATION OF SPONTANEOUS SUSTAINED VENTRICULAR-TACHYCARDIA
    ANDERSON, KP
    WALKER, R
    DUSTMAN, TJ
    GREENBERG, GM
    FULLER, MS
    CIRCULATION, 1993, 88 (04) : 118 - 118
  • [22] TREATMENT OF VENTRICULAR TACHYARRHYTHMIAS IN PATIENTS WITH IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS AND ANTITACHYCARDIA PACEMAKERS
    WINTERS, SL
    CAMUNAS, J
    ERGIN, A
    ESTIOKO, M
    GOMES, JA
    CURRENT OPINION IN CARDIOLOGY, 1989, 4 (04) : 555 - 558
  • [23] SUSTAINED VENTRICULAR-TACHYCARDIA - EFFICACY OF DISOPYRAMIDE
    LERMAN, BB
    WAXMAN, HL
    BUXTON, AE
    JOSEPHSON, ME
    CLINICAL RESEARCH, 1982, 30 (02): : A200 - A200
  • [24] TERMINATION OF SUSTAINED VENTRICULAR-TACHYCARDIA WITH A NEW ANTITACHYCARDIA PACEMAKER - ROLE OF THE NONAUTOMATIC MODE TO FOLLOW PACEMAKER FUNCTION
    FROMER, M
    KUS, T
    PAGE, P
    SHENASA, M
    CANADIAN JOURNAL OF CARDIOLOGY, 1989, 5 (01) : 42 - 46
  • [25] SUSTAINED VENTRICULAR-TACHYCARDIA ASSOCIATED WITH SARCOIDOSIS - ASSESSMENT OF THE UNDERLYING CARDIAC ANATOMY AND THE PROSPECTIVE UTILITY OF PROGRAMMED VENTRICULAR STIMULATION, DRUG-THERAPY AND AN IMPLANTABLE ANTITACHYCARDIA DEVICE
    WINTERS, SL
    COHEN, M
    GREENBERG, S
    STEIN, B
    CURWIN, J
    PE, E
    GOMES, JA
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1991, 18 (04) : 937 - 943
  • [26] IMPLANTED AUTOMATIC BURST PACEMAKERS FOR TERMINATION OF VENTRICULAR-TACHYCARDIA
    FISHER, JD
    FURMAN, S
    KIM, SG
    AMERICAN JOURNAL OF CARDIOLOGY, 1980, 45 (02): : 458 - 458
  • [27] ROLE OF IMPLANTABLE PACEMAKERS IN CONTROL OF RECURRENT VENTRICULAR-TACHYCARDIA
    FISHER, JD
    KIM, SG
    FURMAN, S
    MATOS, JA
    AMERICAN JOURNAL OF CARDIOLOGY, 1982, 49 (01): : 194 - 206
  • [28] FRAGMENTED VENTRICULAR POTENTIALS IN PATIENTS WITH SUSTAINED VENTRICULAR-TACHYCARDIA
    BAILEY, BP
    TAN, A
    BAIRD, DK
    AUSTRALIAN AND NEW ZEALAND JOURNAL OF MEDICINE, 1982, 12 (03): : 324 - 325
  • [29] DISCRIMINATION OF VENTRICULAR-TACHYCARDIA FROM SINUS TACHYCARDIA BY ANTITACHYCARDIA DEVICES - VALUE OF MEDIAN FILTERING
    CHIANG, CMJ
    JENKINS, JM
    DICARLO, LA
    MEDICAL ENGINEERING & PHYSICS, 1994, 16 (06) : 513 - 517
  • [30] RANDOMIZED CONTROLLED COMPARISON OF ANTITACHYCARDIA PACING ALGORITHMS FOR TERMINATION OF VENTRICULAR-TACHYCARDIA
    NEWMAN, D
    DORIAN, P
    HARDY, J
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1993, 21 (06) : 1413 - 1418