CHANGES IN OSCILLOMETRIC PULSE AMPLITUDE ENVELOPE WITH CUFF SIZE - IMPLICATIONS FOR BLOOD-PRESSURE MEASUREMENT CRITERIA AND CUFF SIZE SELECTION

被引:16
|
作者
NG, KG
SMALL, CF
机构
[1] Department of Mechanical Engineering, Queen's University, Kingston, Ont. K7L 3N6, McLaughlin Hall
来源
JOURNAL OF BIOMEDICAL ENGINEERING | 1993年 / 15卷 / 04期
关键词
BLOOD PRESSURE; NONINVASIVE MEASUREMENT; OSCILLOMETRY; CUFF;
D O I
10.1016/0141-5425(93)90003-H
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Oscillometric blood pressures are derived from the amplitude envelope of oscillometric pulses generated in an occlusive cuff during cuff inflation or deflation; one factor which will affect the characteristics of these pulses is the size of the cuff bladder. Because limiting values are stipulated in recommendations and standards for bladder sizes, there is a wide variety of acceptable cuff sizes for any particular application. An experimental and theoretical study was undertaken to show the dependence of oscillometric blood pressures on bladder size. Actual cuff-arm compliance data were obtained from two subjects for two cuffs of different bladder size. Theoretical analysis was then applied to the data to predict the effects of different bladder sizes on the characteristics of the pulses. The results show that cuff-arm compliance and bladder size interact to affect the pulse amplitude and hence oscillometric blood pressure determination. These results suggest that blood pressures obtained using the oscillometric method may vary depending on cuff size, and in particular that replacement cuffs for oscillometric non-invasive blood pressure monitors should be chosen carefully.
引用
收藏
页码:279 / 282
页数:4
相关论文
共 50 条
  • [31] THE EFFECT OF ARM CIRCUMFERENCE AND BLOOD PRESSURE CUFF SIZE ON BLOOD PRESSURE MEASUREMENT IN PREGNANCY
    Mangos, G. J.
    Ong, S.
    Kho, L.
    Brown, M.
    NEPHROLOGY, 2008, 13 : A129 - A129
  • [32] IMPACT OF USING INAPPROPRIATE CUFF SIZE IN BLOOD PRESSURE CLASSIFICATION: ANALYSIS FOR AUSCULTATORY AND OSCILLOMETRIC METHODS
    Albuquerque, Nila
    Sheppard, James
    Araujo, Thelma
    JOURNAL OF HYPERTENSION, 2021, 39 : E129 - E130
  • [33] EFFECT OF SIZE OF CUFF BLADDER ON ACCURACY OF MEASUREMENT OF INDIRECT BLOOD PRESSURE
    SIMPSON, JA
    JAMIESON, G
    DICKHAUS, DW
    GROVER, RF
    AMERICAN HEART JOURNAL, 1965, 70 (02) : 208 - &
  • [34] Cuff size influences blood pressure measurement in obese children and adolescents
    Muhamed, Parosh Kadir
    Olsen, Michael Hecht
    Holm, Jens-Christian
    Ibsen, Hans
    Hvidt, Kristian Nebelin
    DANISH MEDICAL JOURNAL, 2016, 63 (01):
  • [35] SIZE OF BLOOD PRESSURE CUFF FOR JAPANESE ADULTS
    NUKADA, A
    ISHIZAKI, T
    NAGATA, H
    TOHOKU JOURNAL OF EXPERIMENTAL MEDICINE, 1963, 79 (03): : 293 - 305
  • [36] Measurement of blood pressure in hospitals: Is the use of a standard cuff size adequate?
    Comparsi, A
    da Costa, BP
    Antonello, I
    Ogando, P
    de Souza, LH
    de Figueiredo, CP
    JOURNAL OF HYPERTENSION, 2004, 22 : S188 - S189
  • [37] Comparison of a wrist cuff oscillometric blood pressure measurement device with an upper arm-cuff oscillometric device in obese subjects
    de Senarclens, Olivier
    Feihl, Francois
    Giusti, Vittorio
    Engelberger, Rolf P.
    Rodieux, Frederique
    Gomez, Patrick
    Liaudet, Lucas
    Waeber, Bernard
    JOURNAL OF HYPERTENSION, 2006, 24 : 173 - 173
  • [38] BLOOD-PRESSURE AND BODY-COMPOSITION IN CHILDREN - IMPORTANCE OF ALLOWING FOR CUFF SIZE
    JENNER, DA
    VANDONGEN, R
    BEILIN, LJ
    JOURNAL OF HUMAN HYPERTENSION, 1992, 6 (05) : 367 - 374
  • [39] COMPARATIVE EFFECTS OF CUFF SIZE AND TIGHTNESS OF FIT ON ACCURACY OF BLOOD-PRESSURE MEASUREMENTS
    BANNER, TE
    GRAVENSTEIN, JS
    JOURNAL OF CLINICAL MONITORING, 1991, 7 (04): : 281 - 284
  • [40] Accuracy of wrist cuff oscillometric blood pressure measurement in obese patients
    Wertli, M.
    Ruckstuhl, N.
    Vetter, W.
    Suter, P. M.
    JOURNAL OF HYPERTENSION, 2007, 25 : S25 - S25