WALL-CORRECTION AND ABSORBED-DOSE CONVERSION FACTORS FOR FRICKE DOSIMETRY - MONTE-CARLO CALCULATIONS AND MEASUREMENTS

被引:31
|
作者
MA, CM
ROGERS, DWO
SHORTT, KR
ROSS, CK
NAHUM, AE
BIELAJEW, AF
机构
[1] ROYAL MARSDEN HOSP,SUTTON,SURREY,ENGLAND
[2] NATL RES COUNCIL CANADA,INST NATL MEASUREMENT STAND,OTTAWA K1A 0R6,ONTARIO,CANADA
关键词
D O I
10.1118/1.597128
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
For megavoltage radiotherapy photon beams, EGS4 Monte Carlo calculations show, and experimental measurements confirm with an accuracy of 0.2%, that glass or quartz-walled vials used in Fricke dosimetry increase the dose in the Fricke solution. This is mainly caused by increased electron scattering from the glass which increases the dose to the Fricke solution. The dose perturbation is shown to vary from nothing in a Co-60 beam up to 2% in a 24-MV beam. For plastic vials of similar shapes, calculations demonstrate that the effect is in the opposite direction and even at high energies it is much less (0.2% to 0.5%).
引用
收藏
页码:283 / 292
页数:10
相关论文
共 41 条