QUANTITATIVE MOTOR UNIT ANALYSIS - THE EFFECT OF SAMPLE-SIZE

被引:15
|
作者
ENGSTROM, JW
OLNEY, RK
机构
[1] Department of Neurology, School of Medicine, University of California, San Francisco, California
关键词
QUANTITATIVE ELECTROMYOGRAPHY; INTERTRIAL VARIABILITY; SAMPLE SIZE; DIAGNOSTIC SENSITIVITY; MYOPATHY;
D O I
10.1002/mus.880150304
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
This study of quantitative electromyography examines the influence of sample size on motor unit action potential (MUAP) tolerance limits, intertrial variability, and diagnostic sensitivity. We recorded 20 randomly selected MUAPs from the biceps muscle twice in 21 normal subjects, and once in 10 patients with myopathy. The 95% tolerance limits for mean total duration in normal subjects progressively narrowed from 6.6 to 14.2 ms for 5 MUAPs to 7.4 to 13.0 ms for 20 MUAPs. The 95% tolerance limits for intertrial variability were +/- 22% for mean total duration of 20 MUAPs. Larger sample size had a greater effect on reducing intertrial variability than on narrowing 95% tolerance limits for amplitude and area. Quantitative EMG results for duration supported the presence of myopathy in 2 of 10 patients with analysis of 5 MUAPs, and 9 patients with analysis of 20 MUAPs. Although analysis of 5 potentials may be adequate for diagnosis occasionally, quantitative analysis of 20 MUAPs narrows tolerance limits, reduces intertrial variability, and improves diagnostic sensitivity.
引用
收藏
页码:277 / 281
页数:5
相关论文
共 50 条
  • [41] SAMPLE-SIZE DETERMINATION
    TIREN, LI
    NUCLEAR INSTRUMENTS & METHODS, 1965, 34 (01): : 179 - &
  • [42] SAMPLE-SIZE DETERMINATION
    YOUDEN, WJ
    ANALYTICAL CHEMISTRY, 1965, 37 (06) : A71 - &
  • [43] SAMPLE-SIZE NOMOGRAMS
    CLARKE, DD
    ANNALS OF INTERNAL MEDICINE, 1983, 99 (06) : 875 - 875
  • [44] SAMPLE-SIZE DETERMINATION
    LUCAS, HL
    QUARTERLY REVIEW OF BIOLOGY, 1965, 40 (03): : 327 - &
  • [45] SAMPLE-SIZE REQUIREMENTS
    LE, CT
    JOURNAL OF CHRONIC DISEASES, 1983, 36 (09): : 663 - 664
  • [46] SAMPLE-SIZE FOR THE CHEMICAL-ANALYSIS OF METALS
    RUBINSHTEIN, EA
    POPKOVA, GN
    INDUSTRIAL LABORATORY, 1986, 52 (04): : 297 - 299
  • [47] POWER AND SAMPLE-SIZE
    CURTIS, CR
    SALMAN, MD
    SHOTT, S
    JOURNAL OF THE AMERICAN VETERINARY MEDICAL ASSOCIATION, 1990, 197 (07) : 838 - 840
  • [48] SAMPLE-SIZE DETERMINATION
    CLARK, V
    PLASTIC AND RECONSTRUCTIVE SURGERY, 1991, 87 (03) : 569 - 573
  • [49] ROLE OF SAMPLE-SIZE IN SAMPLE EVALUATION
    BARHILLEL, M
    ORGANIZATIONAL BEHAVIOR AND HUMAN PERFORMANCE, 1979, 24 (02): : 245 - 257
  • [50] OVERESTIMATION OF TEST-PERFORMANCE BY ROC ANALYSIS - EFFECT OF SMALL SAMPLE-SIZE
    SEELEY, GW
    BORGSTROM, MC
    PATTON, DD
    MYERS, KJ
    BARRETT, HH
    JOURNAL OF NUCLEAR MEDICINE, 1984, 25 (05) : P107 - P107