DETERMINATION OF TRACES OF NP-237 IN ENRICHED URANIUM SOLUTIONS USING INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY

被引:11
|
作者
RIGLET, C
PROVITINA, O
DAUTHERIBES, JL
REVY, D
机构
[1] Commissariat à l'Energie Atomique, Centre d'Études de Cadarache, DSD/SEP/SAED
关键词
INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY; ALPHA-SPECTROMETRY; NEPTUNIUM; URANIUM;
D O I
10.1039/ja9920700923
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A procedure using inductively coupled plasma mass spectrometry (ICP-MS) has been developed for determining traces of Np-237 in isotopically enriched solutions of U. The chemical treatment of the samples consists of: extracting Np-237 in a 1 mol dm-3 solutions of nitric acid using 0.5 mol dm-3 thenoyltrifluoroacetone in xylene, leaving 99% of the U matrix in the aqueous phase; re-extracting Np-237 in 10 mol dm-3 nitric acid; heating the extracted solution to white fumes in order to remove the excess of acid; and diluting the resulting solution with 2% HNO3. Optimization of the analytical parameters involved suppressing the spectral overlap between Np-237 and U-238. By applying an additive continuous tension to the quadrupole on m/z 237 only, the interfering part of the U-238 peak is eliminated and good resolution can be achieved between m/z 237 and 238, if the concentration of U is less than 100 mg dm-3. The operating conditions adopted result from a compromise between resolution and sensitivity. The suppression of the analyte signal, arising from U at a high concentration, is corrected for by adding an internal standard to the calibration and measuring solutions. The detection limit is about 0.005 mug of Np per gram of U, for an initial sample of 1 g of U and an additive continuous tension of 7.5 V. The analytical performance of ICP-MS is compared with that of alpha-spectrometry The radiochemical determination of Np-237 involves a tedious pre-treatment and measurement procedure. In contrast, ICP-MS appears to be a good alternative to alpha-spectrometry (for the determination of long-lived radionuclides) because of its simplicity, accuracy, excellent detection limits and rapidity of analysis.
引用
收藏
页码:923 / 927
页数:5
相关论文
共 50 条
  • [21] ANALYSIS OF STEELS USING INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY
    VAUGHAN, MA
    HORLICK, G
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1989, 4 (01) : 45 - 50
  • [22] DETERMINATION OF IODINE AND BROMINE IN PLASMA AND URINE BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY
    ALLAIN, P
    MAURAS, Y
    DOUGE, C
    JAUNAULT, L
    DELAPORTE, T
    BEAUGRAND, C
    ANALYST, 1990, 115 (06) : 813 - 815
  • [23] OPTIMIZATION IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY
    LONG, SE
    BROWN, RM
    ANALYST, 1986, 111 (08) : 901 - 906
  • [24] INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY IN GEOCHEMISTRY
    SIEWERS, U
    MIKROCHIMICA ACTA, 1989, 3 (3-6) : 365 - 372
  • [26] DETERMINATION OF LITHIUM IN BIOLOGICAL SAMPLES BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY
    VANHOE, H
    VANDECASTEELE, C
    VERSIECK, J
    DAMS, R
    ANALYTICA CHIMICA ACTA, 1991, 244 (02) : 259 - 267
  • [27] PRECISE DETERMINATION OF BORON IN TITANIUM BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY
    VANHAECKE, F
    VANHOE, H
    VANDECASTEELE, C
    DAMS, R
    ANALYTICA CHIMICA ACTA, 1991, 244 (01) : 115 - 122
  • [28] Rapid and Accurate Characterization of Enriched Uranium Materials by Inductively Coupled Plasma Mass Spectrometry
    Wang W.
    Xu J.
    Xi R.-Y.
    Wang Y.-L.
    Guo S.-Q.
    Fang S.
    Li Z.-M.
    Lin J.-F.
    Zhang H.-T.
    Zhang X.-L.
    Wang C.-J.
    He Y.-J.
    Journal of Chinese Mass Spectrometry Society, 2023, 44 (06) : 797 - 806
  • [29] Rapid determination of 237Np in soil samples by multi-collector inductively-coupled plasma mass spectrometry and gamma spectrometry
    Xiaowei Yi
    Yanmei Shi
    Jiang Xu
    Xiaobing He
    Haitao Zhang
    Jianfeng Lin
    Journal of Radioanalytical and Nuclear Chemistry, 2013, 298 : 1757 - 1761
  • [30] Rapid determination of 237Np in soil samples by multi-collector inductively-coupled plasma mass spectrometry and gamma spectrometry
    Yi, Xiaowei
    Shi, Yanmei
    Xu, Jiang
    He, Xiaobing
    Zhang, Haitao
    Lin, Jianfeng
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2013, 298 (03) : 1757 - 1761