MEASURING HYDROGEN BY COLD-NEUTRON PROMPT-GAMMA ACTIVATION-ANALYSIS

被引:29
|
作者
LINDSTROM, RM [1 ]
PAUL, RL [1 ]
VINCENT, DH [1 ]
GREENBERG, RR [1 ]
机构
[1] UNIV MICHIGAN,DEPT NUCL ENGN,ANN ARBOR,MI 48109
关键词
D O I
10.1007/BF02035915
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
By irradiating with cold neutrons and avoiding hydrogenous materials of construction, we have developed a PGAA instrument at the Cold Neutron Research Facility at NIST with hydrogen detection limits in the microgram range in many materials. Quantities of 5-10 mug H/g are presently measurable in gram-sized samples of silicon or quartz, and of order 0.01 wt % can be quantitatively measured in complex silicate rocks.
引用
收藏
页码:271 / 275
页数:5
相关论文
共 50 条