STUDY OF A LOW-POWERED HE MICROWAVE-INDUCED PLASMA FOR DETERMINATION OF P, CL, BR, AND I BY ATOMIC-EMISSION SPECTROMETRY

被引:14
|
作者
JIN, QH [1 ]
ZHANG, HQ [1 ]
YE, DM [1 ]
ZHANG, JS [1 ]
机构
[1] FUSHUN PETR INST,DEPT APPL CHEM,FUSHUN,PEOPLES R CHINA
关键词
D O I
10.1006/mchj.1993.1044
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Aqueous sample solutions were nebulized by an ultrasonic nebulizer and the solvent was removed by a desolvation apparatus before introduction of the aerosol into the microwave-induced plasma (MIP). The desolvation system consists of a heating tube and a condenser associated with a concentrated H2SO4 desiccator. The water vapor was removed almost completely with the desolvation system. The detectability of microwave-induced plasma atomic emission spectrometry was therefore improved. The detection limits for P, Cl, Br, and I were 0.0045, 0.12, 0.23. and 0.06 μg/ml, respectively. Based on a study of the emission spectra of P, Cl, Br, and I in the range of 200 to 600 nm in He MIP, the optimum analytical lines for determination of corresponding elements were chosen. © 1993 Academic Press. All rights reserved.
引用
收藏
页码:278 / 286