SERUM ANALYSIS FOR POTASSIUM-IONS USING A FIBER OPTIC SENSOR

被引:17
|
作者
GARCIA, RP
MORENO, FA
GARCIA, MED
SANZMEDEL, A
NARAYANASWAMY, R
机构
[1] UNIV OVIEDO, DEPT PHYS & ANALYT CHEM, C JULIAN CLAVERIA, E-33006 OVIEDO, SPAIN
[2] UNIV MANCHESTER, INST SCI & TECHNOL, DEPT INSTRUMENTAT & ANALYT SCI, MANCHESTER M60 1QD, LANCS, ENGLAND
关键词
POTASSIUM; FIBER OPTIC SENSORS; CHROMOGENIC CROWN ETHERS; TAKAGI REAGENT; HUMAN SERUM;
D O I
10.1016/0009-8981(92)90148-J
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
A fibre optic potassium ion sensor has been developed based on reflectance measurements of the pH-dependent chromogenic crown ether 'Takagi reagent' immobilized on the non-ionic polymeric resin Amberlite XAD2. The sensor response is in the 1.5-100 mM range with a detection limit of 0.25 mM K+. The selectivity factor over sodium is 20. The proposed sensor works with a reproducibility of +/- 0.8% at the 5 mM K+ level. Advantageous design features include ease of construction, simplicity of use, reversibility, short response time (approximately 1 min) and an optimum working pH range (7.0-8.0) for biological fluids. This optical sensor was applied successfully to the direct determination of potassium in human serum. Aqueous potassium standards containing 0. 13 M NaCl and a buffer of pH 8 were used for calibration.
引用
收藏
页码:31 / 40
页数:10
相关论文
共 50 条