ASYMMETRIC RESPONSE PROFILE OF A SCANNING FABRY-PEROT-INTERFEROMETER

被引:28
|
作者
LI, ZY [1 ]
STEDMAN, GE [1 ]
BILGER, HR [1 ]
机构
[1] OKLAHOMA STATE UNIV,SCH ELECT & COMP ENGN,STILLWATER,OK 74078
关键词
9;
D O I
10.1016/0030-4018(93)90586-T
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The finite scanning speed of a scanning Fabry-Perot interferometer induces asymmetry in the response profile to a monochromatic input beam. The amplitude of this profile is mathematically that of a fixed interferometer subjected to an input beam with suitable frequency scanning, and is related to the complex error function. The extent of the asymmetry thus depends on a single parameter, eta=betat(c)2, where beta is the frequency scanning rate and t(c) is the power decay time of the interferometer. To obtain an asymmetry less than 5% in the profile obtained from scanning a monochromatic beam requires a scanning time as much as 100 times of the decay time of the interferometer. Experimental data are in good agreement with our calculations.
引用
收藏
页码:240 / 246
页数:7
相关论文