A KNUDSEN CELL MONOPOL MASS-SPECTROMETER FOR USE IN HIGH-TEMPERATURE THERMODYNAMICS

被引:10
|
作者
VRESTAL, J
TOMISKA, J
机构
[1] MASARYK UNIV, INST THEORET & PHYS CHEM, 61137 BRNO, CZECHOSLOVAKIA
[2] UNIV VIENNA, INST PHYS CHEM, A-1090 VIENNA, AUSTRIA
来源
MONATSHEFTE FUR CHEMIE | 1993年 / 124卷 / 11-12期
关键词
MASS SPECTROMETRY; THERMODYNAMICS; INSTRUMENTATION; KNUDSEN TECHNIQUE;
D O I
10.1007/BF00810017
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
A high temperature Knudsen cell was combined with a monopol mass spectrometer model MX 7304 with the ion source 1560014-T01 (Electron, Ukraine) despite the fact that these apparates have been designed for application in organic chemistry only. Relatively simple engineering details, as well as an extremely careful electrical optimisation of the operating conditions yielded a powerful measuring device for application in high temperature thermodynamics. Test measurements as well as the determination of the second law heats of sublimation of solid Fc, Co, Ni, Cu and Cr proved the efficiency of the new device.
引用
收藏
页码:1099 / 1106
页数:8
相关论文
共 50 条