PHASE-SHIFTING INTERFEROMETRY IN THE PRESENCE OF VIBRATION - A NEW ALGORITHM AND SYSTEM

被引:77
|
作者
WIZINOWICH, PL
机构
[1] University of Arizona, Steward Observatory, Tucson, AZ
来源
APPLIED OPTICS | 1990年 / 29卷 / 22期
关键词
D O I
10.1364/AO.29.003271
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
A modification to the usual phase shifting interferometry algorithm permits measurements to be taken fast enough to essentially freeze out vibrations. Only two interferograms are time critical in this 2 + 1 algorithm; the third is null. The implemented system acquires the two time critical interferograms with a 1-millisecond separation on either side of the interline transfer of a standard CCD video camera, resulting in a reduction in sensitivity to vibration of 1-2 orders of magnitude. The required phase shift is achieved via frequency shifting. Laboratory tests comparing this system with a commercial phase shifting package reveal comparable rms errors when vibrations are low; as expected from an analysis of potential phase errors. However, the 2 + 1 system also succeeded when vibrations were large enough to wash out video rate fringes. © 1990 Optical Society of America.
引用
收藏
页码:3271 / 3279
页数:9