A COMPARISON OF THE STRESS-RESPONSE TO LARYNGOSCOPY - THE MACINTOSH VERSUS THE MCCOY BLADE

被引:56
|
作者
MCCOY, EP
MIRAKHUR, RK
MCCLOSKEY, BV
机构
[1] QUEENS UNIV BELFAST,DEPT ANAESTHET,BELFAST BT9 7BL,ANTRIM,NORTH IRELAND
[2] ROYAL VICTORIA HOSP,BELFAST BT12 6BA,ANTRIM,NORTH IRELAND
关键词
EQUIPMENT; LARYNGOSCOPE; INTUBATION; TRACHAEL; COMPLICATIONS; SYMPATHETIC NERVOUS SYSTEM; CATECHOLAMINES;
D O I
10.1111/j.1365-2044.1995.tb05924.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
The cardiovascular changes and catecholamine concentrations were compared in 20 patients before and after laryngoscopy with either the Macintosh or the McCoy laryngoscope blades. Following induction with fentanyl and thiopentone and muscle relaxation with vecuronium the vocal cards were visualised for 10s with either the Macintosh or McCoy blade, chosen at random. Heart rate, arterial blood pressure and plasma noradrenaline and adrenaline concentrations, were measured before, and at, laryngoscopy, and 1, 3 and 5 min later. There was a significan t increase in both heart rate (33%) and arterial blood pressure (27%) after laryngoscopy using the Macintosh blade (p < 0.01). Use of the McCoy blade did not result in any significant change in either heart rate or arterial blood pressure. There was a slight increase (p > 0.05) in plasma noradrenaline concentrations using the Macintosh blade which persisted for 5 min after laryngoscopy. This was not observed with the McCoy blade. There was no change in plasma adrenaline concentrations with either blade during the 5 min after laryngoscopy when compared to the values before laryngoscopy. It is concluded that the stress response to laryngoscopy is less marked with the use of the McCoy blade and is probably due to a reduction in the force necessary to obtain a clear view of the larynx.
引用
收藏
页码:943 / 946
页数:4
相关论文
共 50 条
  • [31] Comparison between Macintosh, Miller and McCoy laryngoscope blade size 2 in paediatric patients - A randomised controlled trial
    Yadav, Pratishtha
    Kundu, Sudeshna Bhar
    Bhattacharjee, Dhurjoti P.
    INDIAN JOURNAL OF ANAESTHESIA, 2019, 63 (01) : 15 - 20
  • [32] Evaluation of Glottic Visualization and Ease of Intubation at Induction of General Anesthesia: A comparison between Macintosh and Mccoy Blade Laryngoscopes
    Abaid-Ur-Rehman
    Azam, Muhammad
    Khan, Zeeshan
    Nadeem, Amjad
    Farooq, Umer
    Khaleeq, Saadia
    PAKISTAN JOURNAL OF MEDICAL & HEALTH SCIENCES, 2021, 15 (09): : 2178 - 2181
  • [33] Clinical comparison of Macintosh vs Glidescope blade tip proximity to neck skin surface during laryngoscopy
    Charters, S.
    Charters, P.
    BRITISH JOURNAL OF ANAESTHESIA, 2011, 107 (05) : 831P - 832P
  • [34] Laryngoscopy via Macintosh Blade versus GlideScope Success Rate and Time for Endotracheal Intubation in Untrained Medical Personnel
    Nouruzi-Sedeh, Parichehr
    Schumann, Mark
    Groeben, Harald
    ANESTHESIOLOGY, 2009, 110 (01) : 32 - 37
  • [35] The learning curve for laryngoscopy: Airtraq versus Macintosh laryngoscopes
    Baciarello, Marco
    Zasa, Michele
    Manferdini, Maria Elena
    Tosi, Michela
    Berti, Marco
    Fanelli, Guido
    JOURNAL OF ANESTHESIA, 2012, 26 (04) : 516 - 524
  • [36] A comparison of the Truview® blade with the Macintosh blade in adult patients
    Barak, M.
    Philipchuck, P.
    Abecassis, P.
    Katz, Y.
    ANAESTHESIA, 2007, 62 (08) : 827 - 831
  • [37] The learning curve for laryngoscopy: Airtraq versus Macintosh laryngoscopes
    Marco Baciarello
    Michele Zasa
    Maria Elena Manferdini
    Michela Tosi
    Marco Berti
    Guido Fanelli
    Journal of Anesthesia, 2012, 26 : 516 - 524
  • [38] Optimal external laryngeal manipulation versus McCoy blade in active position in patients with poor view of glottis on direct laryngoscopy
    Vasudevan, Arumugam
    Venkat, Ranjani
    Badhe, Ashok Shankar
    INDIAN JOURNAL OF ANAESTHESIA, 2010, 54 (01) : 45 - 48
  • [39] Truview™ laryngoscope - A comparison with the macintosh blade
    Harvey, K. E.
    Davies, R. G.
    Hall, J. E.
    Mecklenburgh, J. S.
    EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2004, 21 : 71 - 71
  • [40] CELLULAR STRESS-RESPONSE
    SANDERS, BM
    DYER, SD
    ENVIRONMENTAL TOXICOLOGY AND CHEMISTRY, 1994, 13 (08) : 1209 - 1210