TANDEM MASS-SPECTROMETRIC ANALYSIS OF PEPTIDES AT THE FEMTOMOLE LEVEL

被引:14
|
作者
KENNY, PTM [1 ]
ORLANDO, R [1 ]
机构
[1] SUNTORY INST BIOORGAN RES,SHIMAMOTO,OSAKA 618,JAPAN
关键词
D O I
10.1021/ac00032a022
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In the sequential analysis of peptides by fast atom bombardment (FAB) tandem mass spectrometry (MS/MS), the principal obstacle to decreasing sample quantities was determined to be the signal-to-background ratio of the ionization/desorption process. By decreasing the background ion current, continuous-flow FAB allows complete analysis (both MS and MS/MS) to be performed on 25-75% less sample than required for a conventional FABMS/MS experiment alone. The combination of CF-FAB with array detection permitted sequential analysis of several peptides (900-2000 Da) at the 900 fmol to 5.8 pmol level, without interference from the background. These levels do not produce a molecular ion species easily discernible above the background in conventional FAB.
引用
收藏
页码:957 / 960
页数:4
相关论文