EFFECT OF PRIMARY COIL SIZE ON EDDY-CURRENT DECAY RESISTIVITY MEASUREMENTS

被引:0
|
作者
HARTWIG, KT [1 ]
HUA, CY [1 ]
MCDONALD, LC [1 ]
机构
[1] GEN RES INST NONFERROUS METALS,BEIJING,PEOPLES R CHINA
关键词
MEASURING METHODS; PHYSICAL PROPERTIES; RESISTIVITY;
D O I
10.1016/0011-2275(91)90168-V
中图分类号
O414.1 [热力学];
学科分类号
摘要
Primary coil dimensions can influence the accuracy of resistivity measurements made by the eddy current decay (ECD) method. Experimental results revealed that when the length of a 22.5 mm i.d. primary coil is changed from 60 to 20 mm, the corresponding measured values of resistivity increase by 4-5%. The effect occurs when there is an insubstantial background field at the sample ends. ECD measurement accuracy is better than 1% when the primary coil length to inside diameter ratio is larger than 2.2 and the sample diameter is 44% or less of the primary coil i.d. The investigation included measurements at room temperature, 77 K and 4.2 K on samples having residual resistivity ratios of 153, 340, 1010 and 7790.
引用
收藏
页码:153 / 158
页数:6