Accuracy and reliability of stitched cone-beam computed tomography images

被引:14
|
作者
Egbert, Nicholas [1 ]
Cagna, David R. [2 ]
Ahuja, Swati [2 ]
Wicks, Russell A. [2 ]
机构
[1] Reconstruct Dent Specialists Utah, Salt Lake City, UT USA
[2] Univ Tennessee, Hlth Sci Ctr, Coll Dent, Dept Prosthodont, Memphis, TN 38163 USA
关键词
Cone-Beam Computed Tomography; Computer-Aided Design; Dental Implants;
D O I
10.5624/isd.2015.45.1.41
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: This study was performed to evaluate the linear distance accuracy and reliability of stitched small field of view (FOV) cone-beam computed tomography (CBCT) reconstructed images for the fabrication of implant surgical guides. Material and Methods: Three gutta percha points were fixed on the inferior border of a cadaveric mandible to serve as control reference points. Ten additional gutta percha points, representing fiduciary markers, were scattered on the buccal and lingual cortices at the level of the proposed complete denture flange. A digital caliper was used to measure the distance between the reference points and fiduciary markers, which represented the anatomic linear dimension. The mandible was scanned using small FOV CBCT, and the images were then reconstructed and stitched using the manufacturer's imaging software. The same measurements were then taken with the CBCT software. Results: The anatomic linear dimension measurements and stitched small FOV CBCT measurements were statistically evaluated for linear accuracy. The mean difference between the anatomic linear dimension measurements and the stitched small FOV CBCT measurements was found to be 0.34 mm with a 95% confidence interval of +0.24 - +0.44 mm and a mean standard deviation of 0.30 mm. The difference between the control and the stitched small FOV CBCT measurements was insignificant within the parameters defined by this study. Conclusion: The proven accuracy of stitched small FOV CBCT data sets may allow image-guided fabrication of implant surgical stents from such data sets.
引用
收藏
页码:41 / 47
页数:7
相关论文
共 50 条
  • [31] Anatomical evaluation of the craniovertebral junction on cone-beam computed tomography images
    Tanrisever, Sefkan
    Orhan, Mustafa
    Bahsi, Ilhan
    Yalcin, Eda Didem
    SURGICAL AND RADIOLOGIC ANATOMY, 2020, 42 (07) : 797 - 815
  • [32] Spatially Consistent Supervoxel Correspondences of Cone-Beam Computed Tomography Images
    Pei, Yuru
    Yi, Yunai
    Ma, Gengyu
    Kim, Tae-Kyun
    Guo, Yuke
    Xu, Tianmin
    Zha, Hongbin
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2018, 37 (10) : 2310 - 2321
  • [33] Anatomical evaluation of the craniovertebral junction on cone-beam computed tomography images
    Sefkan Tanrisever
    Mustafa Orhan
    İlhan Bahşi
    Eda Didem Yalçin
    Surgical and Radiologic Anatomy, 2020, 42 : 797 - 815
  • [34] Evaluation of transverse maxillary growth on cone-beam computed tomography images
    Seubert, Brandon J.
    Gaalaas, Laurence
    Larson, Brent E.
    Grunheid, Thorsten
    SCIENTIFIC REPORTS, 2021, 11 (01)
  • [35] Morphometric Analysis of the Sella Turcica on Cone-Beam Computed Tomography Images
    Ince, Rumeysa
    Cihan, Omer F.
    Bahsi, Ilhan
    Yalcin, Eda D.
    JOURNAL OF CRANIOFACIAL SURGERY, 2024, 35 (07) : 1921 - 1925
  • [36] Evaluation of transverse maxillary growth on cone-beam computed tomography images
    Brandon J. Seubert
    Laurence Gaalaas
    Brent E. Larson
    Thorsten Grünheid
    Scientific Reports, 11
  • [37] Reliability and the smallest detectable difference of measurements on 3-dimensional cone-beam computed tomography images
    Damstra, Janalt
    Fourie, Zacharias
    Slater, James J. R. Huddleston
    Ren, Yijin
    AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2011, 140 (03) : E107 - E114
  • [38] Kilovoltage cone-beam computed tomography imaging dose estimation and optimization: Need of daily cone-beam computed tomography
    Trivedi, Gaurav
    Dixit, Chandra K.
    Oinam, Arun S.
    Kapoor, Rakesh
    Bahl, Amit
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2019, 15 (03) : 470 - 474
  • [39] Accuracy and reliability of tooth and root lengths measured on cone-beam computed tomographs
    Sherrard, John F.
    Rossouw, P. Emile
    Benson, Byron W.
    Carrillo, Roberto
    Buschang, Peter H.
    AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2010, 137 (04) : S100 - S108
  • [40] Accuracy and reliability of linear cephalometric measurements from cone-beam computed tomography scans of a dry human skull
    Berco, Mauricio
    Rigali, Paul H., Jr.
    Miner, R. Matthew
    DeLuca, Stephelynn
    Anderson, Nina K.
    Will, Leslie A.
    AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2009, 136 (01) : 17.e1 - 17.e9