FLOW REVERSAL THROUGH A MARK-III HALOTHANE VAPORIZER

被引:1
|
作者
GREGG, AS
JONES, RS
SNOWDON, SL
机构
[1] Department of Anaesthesia, Royal Liverpool University Hospital, Liverpool L69 3BX, Prescot Street
关键词
EQUIPMENT; VAPORIZER; REVERSE FLOW;
D O I
10.1093/bja/71.2.303
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
A fault in the assembly of a Matrix Large Animal Circle anaesthetic machine resulted in reversal of fresh gas flow through the vaporizer. The fault was discovered only after the sudden development of excessive depth of anaesthesia in two equine patients. Laboratory investigations were conducted to determine the effect of flow reversal on vaporizer output. Results indicated that output concentration was approximately doubled under these conditions.
引用
收藏
页码:303 / 304
页数:2
相关论文
共 50 条
  • [1] GAS LEAK THROUGH A FLUOTEC MARK-III VAPORIZER
    ROSENBERG, M
    SOLOD, E
    BOURKE, DL
    ANESTHESIA AND ANALGESIA, 1979, 58 (03): : 239 - 240
  • [2] THE MARK-III
    不详
    ELECTRONICS & WIRELESS WORLD, 1985, 91 (1592): : 24 - 24
  • [3] THE MARK-III SLEEPING CARS
    HILL, RF
    GARRATT, CH
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS, 1982, 196 (MAR): : 87 - 94
  • [4] THE MARK-III VERTEX CHAMBER
    ADLER, J
    BOLTON, T
    BUNNELL, K
    CASSELL, R
    CHEU, E
    EISELE, R
    FREESE, T
    GRAB, C
    MAZAHERI, G
    MIR, R
    ODIAN, A
    PITMAN, D
    STOCKHAUSEN, W
    TOKI, W
    VILLA, F
    WADLEY, W
    WASSERBAECH, S
    WISNIEWSKI, W
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1989, 276 (1-2): : 42 - 52
  • [5] THE ANHEDONIA HYPOTHESIS - MARK-III
    WISE, RA
    BEHAVIORAL AND BRAIN SCIENCES, 1985, 8 (01) : 178 - 184
  • [6] ADVANTAGES OF THE MARK-III CONTAINMENT
    ZACCARIA, A
    NUCLEAR ENGINEERING INTERNATIONAL, 1980, 25 (304): : 49 - 50
  • [7] THE RCAMC ELECTROMYOGRAPH MARK-III
    JASPER, HH
    FORDE, WO
    CANADIAN JOURNAL OF RESEARCH SECTION E-MEDICAL SCIENCES, 1947, 25 (02): : 100 - &
  • [8] MARK-III ROTAHALER PERFORMANCE
    PEARSON, MW
    PRACTITIONER, 1984, 228 (1389) : 251 - 251
  • [9] THE MARK-III STELLAR INTERFEROMETER
    SHAO, M
    COLAVITA, MM
    HINES, BE
    STAELIN, DH
    HUTTER, DJ
    JOHNSTON, KJ
    MOZURKEWICH, D
    SIMON, RS
    HERSHEY, JL
    HUGHES, JA
    KAPLAN, GH
    ASTRONOMY & ASTROPHYSICS, 1988, 193 (1-2) : 357 - 371
  • [10] THE MARK-III NANCAY RADIOHELIOGRAPH
    不详
    SOLAR PHYSICS, 1983, 88 (1-2) : 383 - 390