DETECTION AND QUANTITATION OF ETHYLENE-OXIDE, 2-CHLOROETHANOL, AND ETHYLENE-GLYCOL WITH CAPILLARY GAS-CHROMATOGRAPHY

被引:8
|
作者
DANIELSON, JW
SNELL, RP
OXBORROW, GS
机构
关键词
D O I
10.1093/chromsci/28.3.97
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A procedure is described for detection and quantitation of ethylene oxide and two major reaction products. Detection of ethylene oxide residuals extracted from alcohol pads (cellulose) and plastic material is illustrated. Quantitation of the three compounds can be obtained with a single gas chromatographic (GC) analysis. A DB-Wax-30N, 0.25 mm i.d. column is used with 1:50 split injections with dimethyl-formamide as the solvent and 1:100 split injections with water as the solvent. A DB-Wax-30W, 0.32 mm i.d. column is used for on-column injections and 1:100 split water injections. Linearity of the three compounds is demonstrated.; A procedure is described for detection and quantitation of ethylene oxide and two major reaction products. Detection of ethylene oxide residuals extracted from alcohol pads (cellulose) and plastic material is illustrated. Quantitation of the three compounds can be obtained with a single gas chromatographic (GC) analysis. A DB-Wax-30N, 0.25 mm i.d. column is used with 1:50 split injections with dimethylformamide as the solvent and 1:100 split injections with water as the solvent. A DB-Wax-30W, 0.32 mm i.d. column is used for on-column injections and 1:100 split water injections. Linearity of the three compounds is demonstrated.
引用
收藏
页码:97 / 101
页数:5
相关论文