CALIBRATION WITH RELATIVISTIC AND LOW-VELOCITY IONS OF A CR-39 NUCLEAR TRACK DETECTOR

被引:18
|
作者
BAIOCCHI, P
CECCHINI, S
DEKHISSI, H
GARUTTI, V
GIACOMELLI, G
GIANI, GG
KATSAVOUNIDIS, E
IORI, G
PATRIZII, L
POPA, V
SERRA, P
TOGO, V
VALDRE, U
VILELA, E
机构
[1] IST NAZL FIS NUCL,BOLOGNA,ITALY
[2] CNR,IST TESRE,BOLOGNA,ITALY
[3] UNIV BOLOGNA,DIPARTIMENTO FIS,BOLOGNA,ITALY
[4] UNIV MOHAMED I,FAC SCI,OUJDA,MOROCCO
[5] CALTECH,PASADENA,CA 91125
[6] INST GRAV & SPACE SCI,R-76900 BUCHAREST,ROMANIA
[7] CRE BOLOGNA,ENEA,BOLOGNA,ITALY
关键词
CR-39; CALIBRATION; REL; RELARIVISTIC IONS; LOW VELOCITY IONS; REDUCED ETCH RATE; SEM; REPLICAS;
D O I
10.1016/1350-4487(95)00057-L
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
We present experimental results on the calibration of the CR39 nuclear track detector, manufactured by the Intercast Europe Co., of Parma (Italy). The calibration was performed with several ions of different kinetic energies: from 50 keV protons to 11.3 A GeV gold ions; beta=v/c ranges from about 4 . 10(-3) to about 1. We find that a single curve of the reduced etch rate p versus Restricted Energy Loss is able to describe all data. Furthermore the data are consistent with about 100% contribution of the nuclear energy loss to the CR39 response. This type of CR39 is used in the MACRO experiment at the Gran Sasso Laboratory.
引用
收藏
页码:145 / 150
页数:6
相关论文
共 50 条
  • [1] Calibration with relativistic and low-velocity ions of a CR39 nuclear track detector
    Cecchini, S
    Dekhissi, H
    Garutti, V
    Giacomelli, G
    Katsavounidis, E
    Mandrioli, G
    MargiottaNeri, AR
    Patrizii, L
    Popa, V
    Serra, P
    Spurio, M
    Togo, V
    Valdre, U
    Vilela, E
    NUOVO CIMENTO DELLA SOCIETA ITALIANA DI FISICA A-NUCLEI PARTICLES AND FIELDS, 1996, 109 (08): : 1119 - 1128
  • [2] Calibration of CR-39 solid state nuclear track detector
    P. Wanabongse
    B. Sola
    J. Jamsangtong
    S. Rattanabussayaporn
    Indian Journal of Physics, 2009, 83 : 833 - 837
  • [3] Calibration of solid state nuclear track detector CR-39
    Ren, GX
    Jing, GR
    HIGH ENERGY PHYSICS AND NUCLEAR PHYSICS-CHINESE EDITION, 2005, 29 (06): : 596 - 598
  • [4] Calibration of CR-39 solid state nuclear track detector
    Wanabongse, P.
    Sola, B.
    Jamsangtong, J.
    Rattanabussayaporn, S.
    INDIAN JOURNAL OF PHYSICS, 2009, 83 (06) : 833 - 837
  • [5] THE LOW VELOCITY RESPONSE OF THE SOLID-STATE NUCLEAR TRACK DETECTOR CR-39
    SNOWDENIFFT, DP
    PRICE, PB
    PHYSICS LETTERS B, 1992, 288 (3-4) : 250 - 253
  • [6] Can track-etch detector CR-39 record low-velocity GUT magnetic monopoles?
    He, YD
    PHYSICAL REVIEW D, 1998, 57 (05): : 3182 - 3185
  • [7] Calibration of solid state nuclear track detector CR-39 with monoenergetic protons
    Duan Xiao-Jiao
    Tan Zhi-Xin
    Lan Xiao-Fei
    Huang Yong-Sheng
    Guo Shi-Lun
    Yang Da-Wei
    Tang Xiu-Zhang
    Wang Nai-Yan
    ACTA PHYSICA SINICA, 2010, 59 (05) : 3147 - 3153
  • [8] Response of nuclear track detector CR-39 to low energy muons
    Thomas, H. S. P.
    Deas, R. M.
    Kirkham, L. N.
    Dodd, P. M.
    Zemaityte, E.
    Hillier, A. D.
    Neely, D.
    PLASMA PHYSICS AND CONTROLLED FUSION, 2021, 63 (12)
  • [9] A parameterization of nuclear track profiles in CR-39 detector
    Azooz, A. A.
    Al-Nia'emi, S. H.
    Al-Jubbori, M. A.
    COMPUTER PHYSICS COMMUNICATIONS, 2012, 183 (11) : 2470 - 2479
  • [10] CR-39 nuclear track detector: An experimental guide
    Rana, Mukhtar Ahmed
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2018, 910 : 121 - 126