LINEAR EXCITATION AND DETECTION IN FOURIER-TRANSFORM ION-CYCLOTRON RESONANCE MASS-SPECTROMETRY

被引:32
|
作者
GROSSHANS, PB [1 ]
CHEN, R [1 ]
LIMBACH, PA [1 ]
MARSHALL, AG [1 ]
机构
[1] OHIO STATE UNIV,DEPT CHEM,COLUMBUS,OH 43210
关键词
CYCLOTRON; DETECTION; EXCITATION; FOURIER TRANSFORM; FTMS; ICR;
D O I
10.1016/0168-1176(94)90034-5
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
We present the first Fourier transform ion cyclotron resonance (FT-ICR) ion trap designed to produce both a linear spatial variation of the excitation electric potential field and a linear response of the detection circuit to the motion of the confined ions. With this trap, the magnitude of the detected signal at a given ion cyclotron frequency varies linearly with both the number of ions of given mass-to-charge ratio and also with the magnitude-mode excitation signal at the ion cyclotron orbital frequency; the proportionality constant is mass independent. Interestingly, this linearization may be achieved with any ion trap geometry. The excitation/detection design consists of an array of capacitively coupled electrodes which provide a voltage-divider network that produces a nearly spatially homogeneous excitation electric field throughout the linearized trap; resistive coupling to the electrodes isolates the a.c. excitation (or detection) circuit from the d.c. (trapping) potential. The design is based on analytical expressions for the potential associated with each electrode, from which we are able to compute the deviation from linearity for a trap with a finite number of elements. Based on direct experimental comparisons to an unmodified cubic trap, the linearized trap demonstrates the following performance advantages at the cost of some additional mechanical complexity: (a) signal response linearly proportional to excitation electric field amplitude; (b) vastly reduced axial excitation/ejection for significantly improved ion relative abundance accuracy; (c) elimination of harmonics and sidebands of the fundamental frequencies of ion motion. As a result, FT-ICR mass spectra are now more reproducible. Moreover, the linearized trap should facilitate the characterization of other fundamental aspects of ion behavior in an ICR ion trap, e.g. effects of space charge, nonquadrupolar electrostatic trapping field, etc. Furthermore, this novel design should improve significantly the precision of ion relative abundance and mass accuracy measurements, while removing spectral artifacts of the detection process. We discuss future modifications that linearize the spatial variation of the electrostatic trapping electric field as well, thereby completing the linearization of the entire FT-ICR mass spectrometric techniques. Suggested FT-ICR mass spectrometric applications for the linearized trap are discussed.
引用
收藏
页码:169 / 189
页数:21
相关论文
共 50 条