EFFECT OF AMBIENT-TEMPERATURE AND HUMIDITY ON PERFORMANCE OF BLOOD-GLUCOSE METERS

被引:27
|
作者
KING, JM [1 ]
EIGENMANN, CA [1 ]
COLAGIURI, S [1 ]
机构
[1] PRINCE WALES HOSP,DEPT ENDOCRINOL,RANDWICK,NSW 2031,AUSTRALIA
关键词
TEMPERATURE; HUMIDITY; BLOOD GLUCOSE METERS;
D O I
10.1111/j.1464-5491.1995.tb00488.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The accuracy of four blood glucose meters (Accutrend, ExacTech Companion, Medisense Companion 2, and Glucometer III) was tested at temperatures ranging from 4 to 44 degrees C (control solutions) and 8 degrees C, 24 degrees C, and 36 degrees C (venous blood) and at humidities of 60 % and 80 %. Low and high temperatures resulted in a number of statistically significant changes in glucose readings with all meters. However, Accutrend, Medisense Companion 2, and Glucometer III were 100 % clinically accurate at all temperatures. With the ExacTech Companion, only 70.8 % of control solution and 55.6 % of venous blood results were clinically accurate. The main errors were: (1) cold temperatures lowered the result so that euglycaemic levels erroneously read in the hypoglycaemic range and hyperglycaemic levels gave a better than actual result and (2) hot temperatures increased the result whereby hypoglycaemic levels falsely gave a euglycaemic result. Weather conditions at which blood glucose meters may be operated can affect results and potentially lead to errors in clinical decisions.
引用
收藏
页码:337 / 340
页数:4
相关论文
共 50 条
  • [1] EFFECT OF AMBIENT-TEMPERATURE AND HUMIDITY ON PERFORMANCE OF BLOOD-GLUCOSE MONITORS
    KING, J
    EIGENMANN, C
    COLAGIURI, S
    DIABETES, 1993, 42 : A25 - A25
  • [2] PERFORMANCE OF BLOOD-GLUCOSE METERS
    ELLIOTT, BA
    LANCET, 1979, 1 (8131): : 1410 - 1410
  • [3] PERFORMANCE OF BLOOD-GLUCOSE METERS
    BORTHWICK, LJ
    ROSS, IS
    LANCET, 1979, 1 (8122): : 924 - 924
  • [4] PERFORMANCE OF HOME BLOOD-GLUCOSE METERS
    DAY, JL
    ARCHER, AG
    METCALFE, J
    LANCET, 1979, 2 (8132): : 39 - 40
  • [5] PERFORMANCE OF HOME BLOOD-GLUCOSE METERS
    ROSS, IS
    BORTHWICK, LJ
    LANCET, 1979, 2 (8136): : 257 - 258
  • [6] MONITORING BLOOD-GLUCOSE METERS
    RAINEY, PM
    JATLOW, P
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1995, 103 (02) : 125 - 126
  • [7] EFFECT OF AMBIENT-TEMPERATURE AND HUMIDITY ON URINE OUTPUT IN SHEEP
    GUERRINI, VH
    KOSTER, N
    BERTCHINGER, H
    AMERICAN JOURNAL OF VETERINARY RESEARCH, 1980, 41 (11) : 1851 - 1853
  • [8] THE EFFECT OF AMBIENT-TEMPERATURE ON GLUCOSE-TOLERANCE
    AKANJI, AO
    OPUTA, RN
    DIABETIC MEDICINE, 1991, 8 (10) : 946 - 948
  • [9] EFFECT OF AMBIENT-TEMPERATURE ON LAMB PERFORMANCE
    BRINK, DR
    AMES, DR
    JOURNAL OF ANIMAL SCIENCE, 1975, 41 (01) : 264 - 264
  • [10] EFFECT OF AMBIENT-TEMPERATURE AND HUMIDITY ON PLASMA-CORTISOL IN SHEEP
    GUERRINI, VH
    BERTCHINGER, H