DETERMINATION OF TRACE LEVELS OF SULFATE IN WATER USING FLOW-INJECTION AND IN-LINE PRECONCENTRATION

被引:20
|
作者
KARLSSON, M
PERSSON, JA
MOLLER, J
机构
[1] Tecator AB, S-263 21 Höganä's
关键词
FLOW SYSTEM; SPECTROPHOTOMETRY; SULFATE; WATERS;
D O I
10.1016/S0003-2670(00)82485-X
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Sulphate is preconcentrated on a strong anion-exchange resin and determined using the effect of sulphate ions on the complexation of methylthymol blue and barium. A computer-controlled flow-injection analyser is used to automate the whole procedure. The resin has two functions: it preconcentrates sulphate and also separates sulphate from divalent cations that may interfere in the determination step. The system can handle 30 samples per hour and has a working range from 25 to 1000-mu-g 1(-1) of sulphate. Lower detection limits can be obtained by changing the preconcentration conditions. The effect of both anionic and cationic interferents was studied.
引用
收藏
页码:109 / 113
页数:5
相关论文