THE SUITABILITY OF THE AUSTRALASIAN REFERENCE THROMBOPLASTIN FOR USE ON AN AUTOMATED CENTRIFUGAL ANALYZER

被引:0
|
作者
RAY, MJ
机构
[1] Department of Haematology, The Prince Charles Hospital, Brisbane, QLD
关键词
INTERNATIONAL SENSITIVITY INDEX; PROTHROMBIN TIME INSTRUMENTATION; AUSTRALASIAN REFERENCE THROMBOPLASTIN;
D O I
10.3109/00313029109063600
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
The Australasian Reference Thromboplastin (ART) is now produced in a low turbidity form (ART-LT) for use in photo-electrical coagulometers. This study was designed to assess the suitability of ART and ART-LT for use on a fully automated centrifugal analyser, the ACL300 Research (ACL). These reagents were compared to Thromborel S (TRL), the reagent currently in successful use in our laboratory. It was found that regular ART was an unsuitable reagent for the ACL due to poor daily quality control testing (QC) precision at an INR of 4.5. The ACL also had the effect of significantly altering the ISI of ART from its recommended value as determined for manual testing. ART-LT was found to be an acceptable reagent on the ACL even though its QC precision was lower than that of TRL. The reagent was stable throughout its 8 wk shelf-life. The ACL significantly altered the ISI (International Sensitivity Index) of ART-LT. This may no longer be a disadvantage, as ART-LT is now provided with an ISI specifically for its use on the ACL.
引用
收藏
页码:335 / 338
页数:4
相关论文
共 50 条
  • [1] THE AUSTRALASIAN REFERENCE THROMBOPLASTIN (ART)
    LAMPOTANG, PRLC
    PATHOLOGY, 1980, 12 (02) : 297 - 297
  • [2] THE AUSTRALASIAN REFERENCE THROMBOPLASTIN (ART) UPDATE ON STANDARDIZATION
    LAMPOTANG, PRL
    PATHOLOGY, 1985, 17 (04) : 672 - 672
  • [3] AUTOMATED ENZYME ASSAYS BY USE OF A CENTRIFUGAL ANALYZER WITH FLUORESCENCE DETECTION
    PEARSON, KW
    SMITH, RE
    MITCHELL, AR
    BISSEL, ER
    CLINICAL CHEMISTRY, 1981, 27 (02) : 256 - 262
  • [4] THE AUSTRALASIAN REFERENCE THROMBOPLASTIN .1. A STUDY IN BLEEDING PATIENTS
    FROST, T
    LAU, KS
    JONES, KOA
    PATHOLOGY, 1981, 13 (03) : 525 - 535
  • [5] THE AUSTRALASIAN REFERENCE THROMBOPLASTIN .2. ARE CORRECTED PROTHROMBIN RATIOS VALID
    FROST, T
    LAU, KS
    JONES, KOA
    PATHOLOGY, 1981, 13 (03) : 537 - 542
  • [6] THE IMPLEMENTATION OF A MULTISTAT CENTRIFUGAL ANALYZER IN AN AUTOMATED LABORATORY
    MENHEERE, PPCA
    HULSMAN, JARJ
    MEIJERS, CAM
    COMPUTER PROGRAMS IN BIOMEDICINE, 1982, 14 (03): : 283 - 290
  • [7] AUTOMATED SIMULTANEOUS COAGULATION TESTING WITH A CENTRIFUGAL ANALYZER
    DUNIKOSKI, LK
    KIEFER, L
    BLOOD, 1977, 50 (05) : 265 - 265
  • [8] SIMULTANEOUS COAGULATION TESTING - ACTIVATED PARTIAL THROMBOPLASTIN TIME ASSAYS WITH A CENTRIFUGAL ANALYZER
    DUNIKOSKI, LK
    KRESS, SC
    KIEFER, L
    ELLIS, J
    CLINICAL CHEMISTRY, 1977, 23 (06) : 1133 - 1134
  • [9] BLEEDING IN ANTI-COAGULATED PATIENTS - A STUDY USING THE AUSTRALASIAN REFERENCE THROMBOPLASTIN
    FROST, T
    LAU, KS
    JONES, KOA
    PATHOLOGY, 1981, 13 (03) : 641 - 642
  • [10] AUTOMATED ENZYMATIC DETERMINATION OF L-LACTATE IN SERUM, WITH USE OF A MINIATURE CENTRIFUGAL ANALYZER
    HADJIIOANNOU, TP
    HADJIIOANNOU, SI
    BRUNK, SD
    MALMSTADT, HV
    CLINICAL CHEMISTRY, 1976, 22 (12) : 2038 - 2041