Validation of Cadmium Determination in Polyethylene Reference Materials by Means of Solid Sampling Zeeman Atomic Absorption Spectrometry

被引:1
|
作者
Detcheva, Albena K. [1 ]
Grobecker, Karl H. [2 ]
机构
[1] Bulgarian Acad Sci, Inst Gen & Inorgan Chem, Acad G Bontchev Str,BI 11, BG-1113 Sofia, Bulgaria
[2] European Commiss Joint Res Ctr, Inst Reference Mat & Measurements, B-2440 Geel, Belgium
来源
EURASIAN JOURNAL OF ANALYTICAL CHEMISTRY | 2008年 / 3卷 / 01期
关键词
Cadmium; Polyethylene; Solid Sampling Zeeman Atomic Absorption Spectrometry; Method Validation; Certified Reference Materials;
D O I
10.12973/ejac.2008.00027a
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Direct solid sampling Zeeman atomic absorption spectrometric (SS-ZAAS) method was developed for validation of cadmium determination in polyethylene. Measurements were performed by a third generation Zeeman atomic absorption spectrometer combined with an automatic solid sampler. Applied on alternative wavelength (326.1 nm) the magnetic-3-fieldmeasurement mode extended the linear range of the calibration up to 2500 ng cadmium. High analyte contents (up to 3500 ng Cd) were detectable without laborious dilution of solid samples. Limit of detection was calculated to be 12 ng cadmium according to the 3s-criterion, the maximum sample weight was about 10 mg. The use of matrix-matched solid certified reference materials (CRMs) allowed simple calibration and modifier-free determination of cadmium in polyethylene resulting in high accuracy of the method.