STUDIES OF WALL SHEAR AND MASS-TRANSFER IN A LARGE-SCALE MODEL OF NEONATAL HIGH-FREQUENCY JET VENTILATION

被引:8
|
作者
MULLER, WJ
GERJARUSEK, S
SCHERER, PW
机构
[1] UNIV PENN,DEPT BIOENGN,220 S 33RD ST,TOWNE BLDG,PHILADELPHIA,PA 19104
[2] MIT,DEPT CHEM ENGN,CAMBRIDGE,MA 02139
关键词
Hot-wire anemometry; Jet location in endotracheal tube; Mass and momentum transfer analogy; Velocity profile;
D O I
10.1007/BF02368418
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The problem of endotracheal erosion associated with neonatal high-frequency jet ventilation (HFJV) is investigated through measurement of air velocity profiles in a scaled up model of the system. Fluid mechanical scaling principles are applied in order to construct a model within which velocity profiles are measured by hot-wire anemometry. The effects of two different jet geometries are investigated. Velocity gradients measured near the tracheal wall are used to measure the shear stresses caused by the jet flow on the wall. The Chilton-Colburn analogy between the transport of momentum and mass is applied to investigate tracheal drying caused by the high shear flow. Shear forces are seen to be more than two times higher for jets located near the endotracheal tube wall than for those located axisymmetrically in the center of the tube. Since water vapor fluxes are dependent on these shears, they are also higher for the asymmetric case. Fluxes are shown to be greatly dependent on the temperature and relative humidity of the inspired gas. Water from the tracheal surface may be depleted within one second if inspired gases are inadequately heated and humidified. It is recommended that the design of neonatal HFJV devices include delivery of heated (near body temperature), humidified (as close to 100% humidity as possible) gases through body temperature), humidified (as close to 100% humidity as possible) gases through an axisymmetric jet to best avoid the problem of endotracheal erosion. © 1990 Pergamon Press plc.
引用
收藏
页码:69 / 88
页数:20
相关论文
共 50 条
  • [1] NEONATAL HIGH-FREQUENCY JET VENTILATION
    POKORA, T
    BING, D
    MAMMEL, M
    BOROS, S
    PEDIATRICS, 1983, 72 (01) : 27 - 32
  • [2] HIGH-FREQUENCY JET VENTILATION IN A NEONATAL FOAL
    BAIN, FT
    BROCK, KA
    KOTERBA, AM
    JOURNAL OF THE AMERICAN VETERINARY MEDICAL ASSOCIATION, 1988, 192 (07) : 920 - 922
  • [3] NEONATAL HIGH-FREQUENCY JET VENTILATION - BENEFITS AND RISKS
    LEWALLEN, PK
    BOROS, SJ
    MAMMEL, MC
    COLEMAN, JM
    OPHOVEN, JP
    CLINICAL RESEARCH, 1985, 33 (01): : A148 - A148
  • [4] NEONATAL HIGH-FREQUENCY JET VENTILATION - 4 YEARS EXPERIENCE
    BOROS, SJ
    MAMMEL, MC
    COLEMAN, JM
    LEWALLEN, PK
    GORDON, MJ
    BING, DR
    OPHOVEN, JP
    PEDIATRICS, 1985, 75 (04) : 657 - 663
  • [5] HIGH-FREQUENCY JET VENTILATION IN NEONATAL PULMONARY-HYPERTENSION
    CARLO, WA
    BEOGLOS, A
    CHATBURN, RL
    WALSH, MC
    MARTIN, RJ
    AMERICAN JOURNAL OF DISEASES OF CHILDREN, 1989, 143 (02): : 233 - 238
  • [6] NEONATAL HIGH-FREQUENCY JET VENTILATION - 4 YEARS EXPERIENCE
    LEWALLEN, P
    BOROS, SJ
    MAMMEL, MC
    COLEMAN, M
    OPHOVEN, JP
    CLINICAL RESEARCH, 1984, 32 (04): : A814 - A814
  • [7] Safety and efficacy of high-frequency jet ventilation in neonatal transport
    Mainali, E. S.
    Greene, C.
    Rozycki, H. J.
    Gutcher, G. R.
    JOURNAL OF PERINATOLOGY, 2007, 27 (10) : 609 - 613
  • [8] Unsolved Mysteries: High-Frequency Jet Ventilation in the Neonatal ICU
    Miller, Andrew G.
    Bartle, Renee M.
    RESPIRATORY CARE, 2020, 65 (11) : 1784 - 1785
  • [9] Safety and efficacy of high-frequency jet ventilation in neonatal transport
    E S Mainali
    C Greene
    H J Rozycki
    G R Gutcher
    Journal of Perinatology, 2007, 27 : 609 - 613
  • [10] CLINICAL AND EXPERIMENTAL STUDIES ON HIGH-FREQUENCY JET VENTILATION
    FREITAG, L
    KASPAREK, R
    KONIETZKO, N
    THORACIC AND CARDIOVASCULAR SURGEON, 1985, 33 : 87 - 87