Impaction grafted bone chip size effect on initial stability in an acetabular model: Mechanical evaluation

被引:8
|
作者
Holton, Colin [1 ]
Bobak, Peter [1 ]
Wilcox, Ruth [2 ]
Jin, Zhongmin [2 ,3 ]
机构
[1] Leeds Gen Infirm, Dept Trauma & Orthopaed, Leeds LS1 3EX, W Yorkshire, England
[2] Univ Leeds, Inst Med & Biol Engn, Sch Mech Engn, Leeds LS2 9JT, W Yorkshire, England
[3] Xi An Jiao Tong Univ, Sch Mech Engn, Shanghai 200240, Peoples R China
基金
英国工程与自然科学研究理事会;
关键词
Impaction bone graft; Acetabular model; Bone chip size;
D O I
10.1016/j.jor.2013.09.004
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction: Acetabular bone defect reconstruction is an increasing problem for surgeons with patients undergoing complex primary or revision total hip replacement surgery. Impaction bone grafting is one technique that has favourable long-term clinical outcome results for patients who undergo this reconstruction method for acetabular bone defects. Creating initial mechanical stability of the impaction bone graft in this technique is known to be the key factor in achieving a favourable implant survival rate. Different sizes of bone chips were used in this technique to investigate if the size of bone chips used affected initial mechanical stability of a reconstructed acetabulum. Methodology: Twenty acetabular models were created in total. Five control models were created with a cemented cup in a normal acetabulum. Then five models in three different groups of bone chip size were constructed. The three groups had an acetabular protrusion defect reconstructed using either; 2-4 mm(3), 10 mm(3) or 20 mm(3) bone chip size for impaction grafting reconstruction. The models underwent compression loading up to 9500 N and displacement within the acetabular model was measured indicating the initial mechanical stability. Results: This study reveals that, although not statistically significant, the largest (20 mm(3)) bone chip size grafted models have an inferior maximum stiffness compared to the medium (10 mm(3)) bone chip size. Interpretations: Our study suggests that 10 mm(3) size of bone chips provide better initial mechanical stability compared to smaller or larger bone chips. We dismissed the previously held opinion that the biggest practically possible graft is best for acetabular bone graft impaction. Copyright (C) 2013, Professor P K Surendran Memorial Education Foundation. Publishing Services by Reed Elsevier India Pvt. Ltd. All rights reserved.
引用
收藏
页码:177 / 181
页数:5
相关论文
共 50 条
  • [1] Larger bone graft size and washing of bone grafts prior to impaction enhances the initial stability of cemented cups - Experiments using a synthetic acetabular model
    Arts, J. J. Chris
    Verdonschot, Nico
    Buma, Pieter
    Schreurs, B. Willem
    ACTA ORTHOPAEDICA, 2006, 77 (02) : 227 - 233
  • [2] INITIAL MECHANICAL STABILITY OF ACETABULAR PROSTHESES
    LITSKY, AS
    POPHAL, SG
    ORTHOPEDICS, 1994, 17 (01) : 53 - 57
  • [3] The initial stability of cemented acetabular cups can be augmented by mixing morsellized bone grafts with tricalciumphosphate/hydroxyapatite particles in bone impaction grafting
    Bolder, SBT
    Verdonschot, N
    Schreurs, BW
    Buma, P
    JOURNAL OF ARTHROPLASTY, 2003, 18 (08): : 1056 - 1063
  • [4] Initial stability of a highly porous titanium cup in an acetabular bone defect model
    Yoshimoto, Kensei
    Nakashima, Yasuharu
    Wakiyama, Miyo
    Hara, Daisuke
    Nakamura, Akihiro
    Iwamoto, Mikio
    JOURNAL OF ORTHOPAEDIC SCIENCE, 2018, 23 (04) : 665 - 670
  • [5] Mid-term results in revision hip arthroplasty with impaction bone grafted cup reconstruction for acetabular defects
    Sebastian Rohe
    Nicoletta Dörr
    Sabrina Böhle
    Georg Matziolis
    Steffen Brodt
    Eric Röhner
    Scientific Reports, 12
  • [6] Mid-term results in revision hip arthroplasty with impaction bone grafted cup reconstruction for acetabular defects
    Rohe, Sebastian
    Doerr, Nicoletta
    Boehle, Sabrina
    Matziolis, Georg
    Brodt, Steffen
    Roehner, Eric
    SCIENTIFIC REPORTS, 2022, 12 (01)
  • [7] A physical model for simulating fusion of impaction-grafted morselized cancellous bone
    Heiner, AD
    Brown, TD
    JOURNAL OF BIOMECHANICS, 2001, 34 (06) : 811 - 814
  • [8] Primary stability of a press-fit cup in combination with impaction grafting in an acetabular defect model
    Schierjott, Ronja A.
    Hettich, Georg
    Baxmann, Marc
    Morosato, Federico
    Cristofolini, Luca
    Grupp, Thomas M.
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2021, 39 (05) : 929 - 940
  • [9] Effects of Eccentric Screwing on the Initial Stability of the Acetabular Cup in Bone of Different Qualities
    Hsu, Jui-Ting
    Huang, Heng-Li
    Tsai, Ming-Tzu
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 4: IMAGE PROCESSING, BIOSIGNAL PROCESSING, MODELLING AND SIMULATION, BIOMECHANICS, 2010, 25 : 231 - 234
  • [10] Effect of impaction energy on dynamic bone strains, fixation strength, and seating of cementless acetabular cups
    Doyle, Ruben
    van Arkel, Richard J.
    Jeffers, Jonathan R. T.
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2019, 37 (11) : 2367 - 2375