COLLISIONAL ACTIVATION WITH RANDOM NOISE IN ION TRAP MASS-SPECTROMETRY

被引:70
|
作者
MCLUCKEY, SA
GOERINGER, DE
GLISH, GL
机构
[1] Analytical Chemistry Division, Oak Ridge National Laboratory, Oak Ridge
关键词
D O I
10.1021/ac00037a026
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Random noise applied to the end caps of a quadrupole ion trap shown to be an effective means for the collisional activation of trapped ions independent of mass/charge ratio and number of ions. This technique Is compared and contrasted with conventional single-frequency collisional activation for the molecular ion of N,N-dimethylaniline, protonated cocaine, the molecular anion of 2,4,6-trinitrotoluene, and doubly protonated neuromedin U-8. Collisional activation with noise tends to produce more extensive fragmentation than the conventional approach due to the fact that product ions are also kinetically excited in the noise experiment. The efficiency of the noise experiment in producing detectable product ions relative to the conventional approach ranges from being equivalent to being a factor of 3 less efficient. Furthermore, discrimination against low mass/charge product ions is apparent in the data from multiply charged biomolecules. Nevertheless, collisional activation with random noise provides a very simple means overcoming problems associated with the dependence of single-frequency collisional activation on mass/charge ratio and the number of ions in the ion trap.
引用
收藏
页码:1455 / 1460
页数:6
相关论文
共 50 条
  • [1] RESPONSE-TIME OF IONS TO COLLISIONAL ACTIVATION IN ION-TRAP MASS-SPECTROMETRY
    PARADISI, C
    TRALDI, P
    VETTORI, U
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 1993, 7 (07) : 690 - 692
  • [2] ION TRAP MASS-SPECTROMETRY
    COOKS, RG
    GLISH, GL
    MCLUCKEY, SA
    KAISER, RE
    CHEMICAL & ENGINEERING NEWS, 1991, 69 (12) : 26 - &
  • [3] ION TRAP MASS-SPECTROMETRY
    MARCH, RE
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1992, 118 : 71 - 135
  • [4] QUADRUPOLE ION TRAP MASS-SPECTROMETRY
    COOKS, RG
    KAISER, RE
    ACCOUNTS OF CHEMICAL RESEARCH, 1990, 23 (07) : 213 - 219
  • [5] PRINCIPLES OF COLLISIONAL ACTIVATION IN ANALYTICAL MASS-SPECTROMETRY
    MCLUCKEY, SA
    JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1992, 3 (06) : 599 - 614
  • [6] STUDIES IN MASS-SPECTROMETRY WITH THE ION TRAP DETECTOR
    EICHELBERGER, JW
    BUDDE, WL
    BIOMEDICAL AND ENVIRONMENTAL MASS SPECTROMETRY, 1987, 14 (08): : 357 - 362
  • [7] QUADRUPOLE ION TRAP MASS-SPECTROMETRY OF FULLERENES
    MORAND, KL
    HOKE, SH
    EBERLIN, MN
    PAYNE, G
    COOKS, RG
    ORGANIC MASS SPECTROMETRY, 1992, 27 (03): : 284 - 288
  • [8] HIGH-PRESSURE COLLISIONAL ACTIVATION MASS-SPECTROMETRY
    BLOM, K
    MUNSON, B
    JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1983, 105 (12) : 3793 - 3799
  • [9] ELECTROSPRAY IONIZATION COMBINED WITH ION TRAP MASS-SPECTROMETRY
    VANBERKEL, GJ
    GLISH, GL
    MCLUCKEY, SA
    ANALYTICAL CHEMISTRY, 1990, 62 (13) : 1284 - 1295
  • [10] COLLISIONAL ACTIVATION DECOMPOSITION OF ACTINOMYCINS USING TANDEM MASS-SPECTROMETRY
    ROBOZ, J
    NIEVES, E
    HOLLAND, JF
    MCCAMISH, M
    SMITH, C
    BIOMEDICAL AND ENVIRONMENTAL MASS SPECTROMETRY, 1988, 16 (1-12): : 67 - 70