Comparison of Spectral Domain Optical Coherence Tomography and Ultrasonic Pachymetry for Assessment of Central Corneal Thickness

被引:1
|
作者
Celebi, Ali Riza Cenk [1 ]
Mirza, G. Ertugrul [2 ]
机构
[1] Nigde State Hosp, Clin Ophthalmol, Nigde, Turkey
[2] Erciyes Univ, Fac Med, Dept Ophthalmol, Kayseri, Turkey
关键词
Spectral domain optical coherence tomography; central corneal thickness; ultrasonic pachymetry;
D O I
10.4274/tjo.94834
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Objectives: The aim of this study was to determine if there is a difference in central corneal thickness (CCT) measurements obtained by Cirrus spectral domain optical coherence tomography (SD-OCT) and ultrasonic pachymetry in healthy individuals. Materials and Methods: The study included 50 healthy consecutively selected individuals without ocular or systemic disease. CCT was first measured using OCT, and then using ultrasonic pachymetry. Results: Mean age of the participants was 31.44 years. Mean CCT measured using SD-OCT was 531.78 mu m versus 535.15 mu m by ultrasonic pachymetry. Mean CCT measurement obtained by Cirrus SD-OCT showed statistically significant difference by approximately 3.37 mu m than the one obtained by ultrasonic pachymetry (t-test, p<0.05); however, Bland-Altman analysis proved that there was high concordance between the measurements. Conclusion: CCT measurements obtained by Cirrus SD-OCT were very similar to those obtained by ultrasonic pachymetry, and as such we think that Cirrus SD-OCT can be used in our present ophthalmology practice to measure OCT.
引用
收藏
页码:259 / 262
页数:4
相关论文
共 50 条
  • [1] Comparison of Central Corneal Thickness with Ultrasound Pachymetry, Noncontact Specular Microscopy and Spectral Domain Optical Coherence Tomography
    Erdur, Sevil Karaman
    Demirci, Goktug
    Dikkaya, Funda
    Kocabora, Mehmet Selim
    Ozsutcu, Mustafa
    SEMINARS IN OPHTHALMOLOGY, 2018, 33 (06) : 782 - 787
  • [2] A clinical comparison of central corneal thickness measurements with Optical Coherence Tomography versus ultrasonic pachymetry
    Bechmann, M
    Thiel, MJ
    Roesen, B
    Ullrich, S
    Ulbig, MW
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1999, 40 (04) : S902 - S902
  • [3] Comparison of ultrasonic pachymetry and Fourier-domain optical coherence tomography for measurement of corneal thickness in dogs with and without corneal disease
    Hoehn, A. L.
    Thomasy, S. M.
    Kass, P. H.
    Horikawa, T.
    Samuel, M.
    Shull, O. R.
    Stewart, K. A.
    Murphy, C. J.
    VETERINARY JOURNAL, 2018, 242 : 59 - 66
  • [4] Comparison between central corneal thickness measurements by ultrasound pachymetry and optical coherence tomography
    Leung, Dexter Y. L.
    Lam, Douglas K. T.
    Yeung, Barry Y. M.
    Lam, Dennis S. C.
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2006, 34 (08): : 751 - 754
  • [5] Accuracy of RTVue Optical Coherence Tomography, Pentacam, and Ultrasonic Pachymetry for the Measurement of Central Corneal Thickness
    Nam, Sang Min
    Im, Chan Young
    Lee, Hyung Keun
    Kim, Eung Kweon
    Kim, Tae-im
    Seo, Kyoung Yul
    OPHTHALMOLOGY, 2010, 117 (11) : 2096 - 2103
  • [6] Comparison of corneal thickness measured with optical coherence tomography, ultrasonic pachymetry, and a scanning slit method
    Thomas, Joseph
    Wang, Jianhua
    Rollins, Andrew M.
    Sturm, Julie
    JOURNAL OF REFRACTIVE SURGERY, 2006, 22 (07) : 671 - 678
  • [7] Corneal and epithelial thickness in keratoconus: A comparison of ultrasonic pachymetry Orbscan II, and optical coherence tomography
    Haque, S
    Simpson, T
    Jones, L
    JOURNAL OF REFRACTIVE SURGERY, 2006, 22 (05) : 486 - 493
  • [8] Comparison of Measurement of Central Corneal Thickness with Spectral Domain Optical Coherence Tomography and Standard Ultrasonic Pachymeter in Premature Infants
    Hekimoglu, Emre
    Erol, Muhammet Kazim
    Toslak, Devrim
    Coban, Deniz Turgut
    Dogan, Berna
    Yucel, Ozgur
    JOURNAL OF OPHTHALMOLOGY, 2015, 2015
  • [9] Corneal and epithelial thickness in keratoconus: A comparison of ultrasonic pachymetry, orbscan and optical coherence tomography measurements
    Haque, S
    Jones, L
    Simpson, T
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2004, 45 : U28 - U28
  • [10] Central Corneal Thickness Measurements With Fourier-Domain Optical Coherence Tomography Versus Ultrasonic Pachymetry and Rotating Scheimpflug Camera
    Ishibazawa, Akihiro
    Igarashi, Sho
    Hanada, Kazuomi
    Nagaoka, Taiji
    Ishiko, Satoshi
    Ito, Haruna
    Yoshida, Akitoshi
    CORNEA, 2011, 30 (06) : 615 - 619