Development and Validation of a Wear Model to Predict Polyethylene Wear in a Total Knee Arthroplasty: A Finite Element Analysis

被引:20
|
作者
Innocenti, Bernardo [1 ]
Labey, Luc [2 ]
Kamali, Amir [3 ]
Pascale, Walter [4 ]
Pianigiani, Silvia [4 ]
机构
[1] Univ Libre Bruxelles, BEAMS Dept Bio Elect & Mech Syst, Ave Roosevelt 50, B-1050 Brussels, Belgium
[2] Katholieke Univ Leuven, Mech Engn Technol TC, B-2440 Geel, Belgium
[3] Smith & Nephew Orthopaed Ltd, Royal Leamington Spa CV31 3HL, England
[4] IRCCS, Ist Ortoped Galeazzi, I-20161 Milan, Italy
来源
LUBRICANTS | 2014年 / 2卷 / 04期
关键词
wear; TKA; validated model; FEA; patello-femoral joint;
D O I
10.3390/lubricants2040193
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
Ultra-high molecular weight polyethylene (UHMWPE) wear in total knee arthroplasty (TKA) components is one of the main reasons of the failure of implants and the consequent necessity of a revision procedure. Experimental wear tests are commonly used to quantify polyethylene wear in an implant, but these procedures are quite expensive and time consuming. On the other hand, numerical models could be used to predict the results of a wear test in less time with less cost. This requires, however, that such a model is not only available, but also validated. Therefore, the aim of this study is to develop and validate a finite element methodology to be used for predicting polyethylene wear in TKAs. Initially, the wear model was calibrated using the results of an experimental roll-on-plane wear test. Afterwards, the developed wear model was applied to predict patello-femoral wear. Finally, the numerical model was validated by comparing the numerically-predicted wear, with experimental results achieving good agreement.
引用
收藏
页码:193 / 205
页数:13
相关论文
共 50 条
  • [1] An elasto-plastic finite element model for polyethylene wear in total hip arthroplasty
    Teoh, SH
    Chan, WH
    Thampuran, R
    JOURNAL OF BIOMECHANICS, 2002, 35 (03) : 323 - 330
  • [2] Prediction of total knee arthroplasty polyethylene wear using the wear index
    Miura, H
    Higaki, H
    Nakanishi, Y
    Mawatari, T
    Moro-oka, T
    Murakami, T
    Iwamoto, Y
    JOURNAL OF ARTHROPLASTY, 2002, 17 (06): : 760 - 766
  • [3] Factors affecting polyethylene wear in total knee arthroplasty
    Kuster, MS
    Stachowiak, GW
    ORTHOPEDICS, 2002, 25 (02) : S235 - S242
  • [4] EFFECT OF POLYETHYLENE QUALITY ON WEAR IN TOTAL KNEE ARTHROPLASTY
    TANNER, MG
    WHITESIDE, LA
    WHITE, SE
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1995, (317) : 83 - 88
  • [5] Polyethylene Wear in Knee Arthroplasty
    Chakravarty, Rajit
    Elmallah, Randa D. K.
    Cherian, Jeffrey Jai
    Kurtz, Steven M.
    Mont, Michael A.
    JOURNAL OF KNEE SURGERY, 2015, 28 (05) : 370 - 375
  • [6] Accurate in vivo measurement of polyethylene wear in total knee arthroplasty
    Hoshino, A
    Fukuoka, Y
    Ishida, A
    JOURNAL OF ARTHROPLASTY, 2002, 17 (04): : 490 - 496
  • [7] Arthroscopy for evaluation of polyethylene wear after total knee arthroplasty
    Kondo, Makoto
    Fujii, Tadashi
    Kitagawa, Hiroshi
    Tsumura, Hiroshi
    Kadoya, Yoshinori
    JOURNAL OF ORTHOPAEDIC SCIENCE, 2008, 13 (05) : 433 - 437
  • [8] Wear in total knee arthroplasty-just a question of polyethylene?
    Kretzer, Jan Philippe
    Reinders, Joern
    Sonntag, Robert
    Hagmann, Sebastien
    Streit, Marcus
    Jeager, Sebastian
    Moradi, Babak
    INTERNATIONAL ORTHOPAEDICS, 2014, 38 (02) : 335 - 340
  • [9] Wear in total knee arthroplasty—just a question of polyethylene?Metal ion release in total knee arthroplasty
    Jan Philippe Kretzer
    Joern Reinders
    Robert Sonntag
    Sebastien Hagmann
    Marcus Streit
    Sebastian Jeager
    Babak Moradi
    International Orthopaedics, 2014, 38 : 335 - 340
  • [10] Polyethylene wear in primary knee arthroplasty
    Bonaspetti, G.
    Di Fabio, S.
    Fenu, R.
    Costa, L.
    Pazzaglia, U. E.
    JOURNAL OF ORTHOPAEDICS AND TRAUMATOLOGY, 2005, 6 (01) : 15 - 20