CRITICAL ASPECTS OF THE FUNCTION OF INSPIRATORY FLOW DRIVEN INHALERS

被引:31
|
作者
OLSSON, B
ASKING, L
机构
关键词
POWDER INHALERS; DRUG DELIVERY; INSPIRATORY FLOW DRIVEN INHALERS; FLOW RATE; AEROSOLIZATION OF DRUG; RESPIRABLE DOSE; DEVICE RESISTANCE; ASTHMA; INSPIRATORY FORCE; PARTICLE SIZE DISTRIBUTION; BUDESONIDE; TURBUHALER;
D O I
10.1089/jam.1994.7.Suppl_1.S-43
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Powder inhalers rely on the patient's inspiratory flow for delivery and aerosolization of the medicament. Different inhalers offer different resistances to inspiration through the device. The flow rate attained by a patient depends on the effort expended and on the air flow resistance of the device; flow rate is thus a device-dependent variable. A comparison between powder inhalers should therefore take their air flow resistances into account. An empirical relation between the peak inspiratory flow rate through a device and the flow resistance of the device has been developed using results from healthy subjects and verified to a limited extent using data from asthmatic patients. The flow resistance was measured as the pressure drop at a fixed flow rate. The empirical relation can be used to predict the change in peak inspiratory flow rate resulting from a change in flow resistance. Matching flow rates, representing equal inspiratory efforts, may thus be calculated for different inhalation devices.