COMPARISON OF CALCULATED AND MEASURED HETEROGENEITY CORRECTION FACTORS FOR I-125, CS-137, AND IR-192 BRACHYTHERAPY SOURCES NEAR LOCALIZED HETEROGENEITIES

被引:86
|
作者
WILLIAMSON, JF [1 ]
PERERA, H [1 ]
LI, ZF [1 ]
LUTZ, WR [1 ]
机构
[1] UNIV ARIZONA,HLTH SCI CTR,DEPT RADIAT ONCOL,TUCSON,AZ 85724
关键词
BRACHYTHERAPY DOSIMETRY; HETEROGENEITY CORRECTIONS; SILICON DIODE MEASUREMENTS; MONTE-CARLO SIMULATION;
D O I
10.1118/1.597088
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The influence of tissue and applicator heterogeneities on brachytherapy dose distributions is not well understood, despite widespread use of shielded applicators in intracavitary therapy. Heterogeneity correction factors (HCF) have been measured using a silicon diode detector arising from bounded heterogeneities consisting of lead, steel, titanium, silver, aluminum, and air cylinders near brachytherapy sources of I-125, Cs-137, and Ir-192. In addition, transverse-axis dose distributions for the three sources in homogeneous water were measured for distances of 0.2 to 16.0 cm. For each point of measurement, relative diode readings were simulated by a Monte Carlo photon transport code utilizing accurate models of the source internal structure, the experimental measure geometry and the source-strength calibration geometry. Comparison of measured and calculated HCF's reveals excellent agreement (1%-3% average) over a wide range of materials, diameters, and thicknesses. In addition, Monte Carlo simulation not only accurately reproduced the relative transverse-axis dose distributions in homogeneous medium, but was able to predict the variation of diode response with photon energy with an accuracy of 3% over the range of 30-662 keV. Our measurements demonstrate that HCF's vary by as much as 60%-100% with distance and heterogeneity diameter for a fixed thickness. Finally, silicon diode measurements of HCF (defined as reading with heterogeneity/reading in homogeneous medium) is shown to lead to errors of 5%-30% for Cs-137 and Ir-192 sources in the presence of high-atomic number shielding materials. This paper concludes, that Monte Carlo simulation is a powerful, convenient and accurate tool for investigating the long-neglected area of brachytherapy heterogeneity corrections.
引用
收藏
页码:209 / 222
页数:14
相关论文
共 21 条
  • [1] COMPARISON OF MEASURED AND CALCULATED DOSE-RATES IN WATER NEAR I-125 AND IR-192 SEEDS
    WILLIAMSON, JF
    MEDICAL PHYSICS, 1991, 18 (04) : 776 - 786
  • [2] DOSIMETRY ON TRANSVERSE AXES OF I-125 AND IR-192 INTERSTITIAL BRACHYTHERAPY SOURCES
    NATH, R
    MEIGOONI, AS
    MELI, JA
    MEDICAL PHYSICS, 1990, 17 (06) : 1032 - 1040
  • [3] Intracavitary brachytherapy for carcinoma of the uterine cervix: Comparison of HDR (Ir-192) and MDR (Cs-137)
    Tanaka E.
    Suzuki O.
    Oh R.-J.
    Takeda T.
    Teshima T.
    Inoue T.
    Inoue T.
    Radiation Medicine, 2006, 24 (1): : 50 - 57
  • [4] A COMPARISON OF RADIAL DOSE FUNCTIONS FOR PD-103, I-125, SM-145, AM-241, YB-169, IR-192, AND CS-137 BRACHYTHERAPY SOURCES
    MEIGOONI, AS
    NATH, R
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1992, 22 (05): : 1125 - 1130
  • [5] ANISOTROPY FUNCTIONS FOR PD-103, I-125, AND IR-192 INTERSTITIAL BRACHYTHERAPY SOURCES
    NATH, R
    MEIGOONI, AS
    MUENCH, P
    MELILLO, A
    MEDICAL PHYSICS, 1993, 20 (05) : 1465 - 1473
  • [6] On the Evaluation of GATE Monte Carlo Toolkit Performance for the Dosimetry of Ir-192 and I-125 Brachytherapy Sources
    Pappas, E.
    Hammoud, R.
    Kagadis, G.
    Sharif, A.
    Bakas, A.
    Papadimitroulas, P.
    Anagnostopoulos, G.
    Kantemiris, I.
    Loudos, G.
    Al Hammadi, N.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [7] Design of a web page with the dosimetric parameters for all the different models of Cs-137 and Ir-192 sources used in brachytherapy
    Pérez-Calatayud, J
    Ballester, F
    Lis, M
    Casal, E
    Granero, D
    Cases, R
    Venselaar, J
    RADIOTHERAPY AND ONCOLOGY, 2004, 71 : S41 - S41
  • [8] Measurement and calculation of the dose at large distances from brachytherapy sources: Cs-137, Ir-192, and Co-60
    Venselaar, JLM
    vanderGiessen, PH
    Dries, WJF
    MEDICAL PHYSICS, 1996, 23 (04) : 537 - 543
  • [9] Evaluation of the Correction Factor Due to the Lack of Full Scatter Conditions in Cs-137 and Ir-192 Brachytherapy Dosimetric Studies
    Ballester, F.
    Granero, D.
    Perez-Calatayud, J.
    Rivard, M.
    Melhus, C.
    Pujades, M.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [10] EVALUATION OF THE CORRECTION FACTOR DUE TO THE LACK OF FULL SCATTER CONDITIONS IN CS-137 AND IR-192 BRACHYTHERAPY DOSIMETRIC STUDIES
    Ballester, F.
    Perez-Calatayud, J.
    Rivard, M.
    Melhus, C.
    Pujades, M.
    Granero, D.
    RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S398 - S398