Electrochemical platform for simultaneous determination of levodopa, acetaminophen and tyrosine using a graphene and ferrocene modified carbon paste electrode

被引:0
|
作者
Kaveh Movlaee
Hadi Beitollahi
Mohammad Reza Ganjali
Parviz Norouzi
机构
[1] University of Tehran,Center of Excellence in Electrochemistry, Faculty of Chemistry
[2] Graduate University of Advanced Technology,Environment Department, Institute of Science and High Technology and Environmental Sciences
[3] Tehran University of Medical Sciences,Biosensor Research Center, Endocrinology & Metabolism Molecular
来源
Microchimica Acta | 2017年 / 184卷
关键词
Modified electrodes; Carbon paste electrode; Ferrocene mediator; Square wave voltammetry; Electrochemical sensor; Nanotechnology; Drug analysis;
D O I
暂无
中图分类号
学科分类号
摘要
A carbon paste electrode (CPE) was modified with graphene and ethyl 2-(4-ferrocenyl-[1,2,3]triazol-1-yl) acetate (EFTA) and used for electrocatalytic oxidation of levodopa. Under optimum conditions and at pH 7.0, the oxidation of levodopa occurs at a potential about 280 mV (vs. Ag/AgCl) using square wave voltammetry (SWV) which is less positive than that of an unmodified CPE. The modified electrode can well resolve the voltammetric peaks of levodopa, acetaminophen and tyrosine. The peak current is linear in the 0.2 μM to 0.4 mM levodopa concentration range. The respective figures for acetaminophen and tyrosine are 485 mV and 745 mV, respectively, and the analytical ranges are from 1.0 μM to 0.15 mM for acetaminophen, and from 5.0 μM to 0.18 mM for tyrosine.
引用
收藏
页码:3281 / 3289