Intraocular pressure control after the implantation of a second Ahmed glaucoma valve

被引:0
|
作者
Jesús Jiménez-Román
Félix Gil-Carrasco
Vital Paulino Costa
Rui Barroso Schimiti
Fabián Lerner
Priscila Rezende Santana
Jose Paulo Cabral Vascocellos
Armando Castillejos-Chévez
Mauricio Turati
Karina Fabre-Miranda
机构
[1] Asociación Para Evitar la Ceguera en México,Glaucoma Department
[2] UNICAMP,Glaucoma Department
[3] Fundación para el estudio del Glaucoma,Glaucoma Department
来源
关键词
Glaucoma drainage device; Ahmed drainage device; Glaucoma; Surgery;
D O I
暂无
中图分类号
学科分类号
摘要
The objective of this study is to evaluate the efficacy and safety of a second Ahmed glaucoma valve (AGV) in eyes with refractory glaucoma that had undergone prior Ahmed device implantation. This multicenter, retrospective study evaluated 58 eyes (58 patients) that underwent a second AGV (model S2-n = 50, model FP7-n = 8) due to uncontrolled IOP under maximal medical therapy. Outcome measures included IOP, visual acuity, number of glaucoma medications, and postoperative complications. Success was defined as IOP <21 mmHg (criterion 1) or 30 % reduction of IOP (criterion 2) with or without hypotensive medications. Persistent hypotony (IOP <5 mmHg after 3 months of follow-up), loss of light perception, and reintervention for IOP control were defined as failure. Mean preoperative IOP and mean IOPs at 12 and 30 months were 27.55 ± 1.16 mmHg (n = 58), 14.45 ± 0.83 mmHg (n = 42), and 14.81 ± 0.87 mmHg (n = 16), respectively. The mean numbers of glaucoma medications preoperatively at 12 and 30 months were 3.17 ± 0.16 (n = 58), 1.81 ± 0.2 (n = 42), and 1.83 ± 0.35 (n = 18), respectively. The reductions in mean IOP and number of medications were statistically significant at all time intervals (P < 0.001). According to criterion 1, Kaplan–Meier survival curves disclosed success rates of 62.9 % at 12 months and 56.6 % at 30 months. According to criterion 2, Kaplan–Meier survival curves disclosed success rates of 43.9 % at 12 months and 32.9 % at 30 months. The most frequent early complication was hypertensive phase (10.3 %) and the most frequent late complication was corneal edema (17.2 %). Second AGV implantation may effectively reduce IOP in eyes with uncontrolled glaucoma, and is associated with relatively few complications.
引用
收藏
页码:347 / 353
页数:6
相关论文
共 50 条
  • [1] Intraocular pressure control after the implantation of a second Ahmed glaucoma valve
    Jimenez-Roman, Jesus
    Gil-Carrasco, Felix
    Costa, Vital Paulino
    Schimiti, Rui Barroso
    Lerner, Fabian
    Santana, Priscila Rezende
    Cabral Vascocellos, Jose Paulo
    Castillejos-Chevez, Armando
    Turati, Mauricio
    Fabre-Miranda, Karina
    INTERNATIONAL OPHTHALMOLOGY, 2016, 36 (03) : 347 - 353
  • [2] Ahmed valve implantation to control intractable glaucoma after phacoemulsification and intraocular lens implantation in a dog
    García-Sánchez, GA
    Whitley, RD
    Brooks, DE
    Trigo, F
    Piñón, A
    VETERINARY OPHTHALMOLOGY, 2005, 8 (02) : 139 - 144
  • [3] Ahmed Glaucoma Valve Implantation to Reduce Intraocular Pressure: Updated Perspectives
    Arikan, Gul
    Dokuz, Uzeyir Gunenc
    CLINICAL OPHTHALMOLOGY, 2023, 17 : 1833 - 1845
  • [4] Evaluation of success after second Ahmed glaucoma valve implantation
    Nilforushan, Naveed
    Yadgari, Maryam
    Jazayeri, Anis Alsadat
    Karimi, Nasser
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2016, 64 (03) : 206 - 210
  • [5] Intraocular Pressure Control after the Implantation of an Ahmed Glaucoma Device in Eyes with a Failed Trabeculectomy
    Schimiti, Rui Barroso
    Abe, Ricardo Yuji
    Vasconcellos, Jose Paulo C.
    Tavares, Carla Melo
    Costa, Vital P.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2015, 56 (07)
  • [6] A comparison of intraocular pressure response and complications after Ahmed glaucoma valve implantation, with and without concurrent vitrectomy
    Chynn, EW
    Sibayan, SA
    Latina, MA
    Netland, PA
    Kini, MM
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1996, 37 (03) : 1180 - 1180
  • [7] Effect of immediate shallow anterior chamber after Ahmed glaucoma valve implantation on intermediate-term intraocular pressure control
    Patil, Trupti Sudhir
    Mani, Baskaran
    Balekudaru, Shantha
    George, Ronnie J.
    Lingam, Vijaya
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2022, 70 (08) : 2915 - 2921
  • [8] Intraoperative testing of opening and closing pressure predicts risk of low intraocular pressure after Ahmed glaucoma valve implantation
    Bochmann, F.
    Kipfer, A.
    Tarantino, J.
    Kaufmann, C.
    Bachmann, L.
    Thiel, M.
    EYE, 2014, 28 (10) : 1184 - 1189
  • [9] Intraoperative testing of opening and closing pressure predicts risk of low intraocular pressure after Ahmed glaucoma valve implantation
    F Bochmann
    A Kipfer
    J Tarantino
    C Kaufmann
    L Bachmann
    M Thiel
    Eye, 2014, 28 : 1184 - 1189
  • [10] Non-intraocular pressure-related revision surgeries after Ahmed glaucoma valve (AGV) implantation in refractory glaucoma
    Asli Kirmaci Kabakci
    Banu Solmaz
    Berna Basarir
    Burcu Kemer Atik
    Tekin Yasar
    International Ophthalmology, 2021, 41 : 3533 - 3538