Development of a novel shrouded impeller pediatric blood pump

被引:8
|
作者
Khan, Talha Irfan [1 ]
Zad, Haris Sheh [1 ]
Lazoglu, Ismail [1 ]
Yalcin, Ozlem [2 ]
机构
[1] Koc Univ, Mfg & Automat Res Ctr, TR-34450 Istanbul, Turkey
[2] Koc Univ, Sch Med, Istanbul, Turkey
关键词
Pediatric ventricular assist device; Computational fluid dynamics; Shrouded impeller; Hydraulic performance; VENTRICULAR ASSIST DEVICE; CHILDREN; SUPPORT; DAMAGE;
D O I
10.1007/s10047-018-1028-3
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The aim of this work was to analyze a shrouded impeller pediatric ventricular assist device (SIP-VAD). This device has distinctive design characteristics and parameter optimizations for minimization of recirculation flow and reduction in high-stress regions that cause blood damage. Computational Fluid Dynamics (CFD) simulations were performed to analyze the optimized design. The bench-top prototype of SIP-VAD was manufactured with biocompatible stainless steel. A study on the hydrodynamic and hemodynamic performance of the SIP-VAD was conducted with predictions from CFD and actual experimentation values, and these results were compared. The CFD analysis yielded a pressure range of 29-90 mmHg corresponding to flow rates of 0.5-3 L/min over 9000-11000 rpm. The predicted value of the normalized index of hemolysis (NIH) was 0.0048 g/100 L. The experimental results with the bench-top prototype showed a pressure rise of 30-105 mmHg for the flow speed of 8000-12000 rpm and flow rate of 0.5-3.5 L/min. The maximum difference between CFD and experimental results was 4 mmHg pressure. In addition, the blood test showed the average NIH level of 0.00674 g/100 L. The results show the feasibility of shrouded impeller design of axial-flow pump for manufacturing the prototype for further animal trials.
引用
收藏
页码:142 / 149
页数:8
相关论文
共 50 条
  • [1] The Progress in the Novel Pediatric Rotary Blood Pump Sputnik Development
    Telyshev, Dmitry
    Denisov, Maxim
    Pugovkin, Alexander
    Selishchev, Sergey
    Nesterenko, Igor
    ARTIFICIAL ORGANS, 2018, 42 (04) : 432 - 443
  • [2] Rapid manufacturing techniques in the development of an axial blood pump impeller
    Chan, WK
    Wong, YW
    Chua, CK
    Lee, CW
    Feng, C
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2003, 217 (H6) : 469 - 475
  • [3] Performance discontinuity of a shrouded centrifugal pump impeller: Part 2 - stabilizing geometrical modifications
    Ginter, F
    Staubli, T
    THIRD EUROPEAN CONFERENCE ON TURBOMACHINERY - VOLS A AND B: FLUID DYNAMICS AND THERMODYNAMICS, 1999, 1999 (1A-1B): : 1039 - 1049
  • [4] Prediction of Leakage Flow in a Shrouded Centrifugal Blood Pump
    Teo, Ji-Bin
    Chan, Weng-Kong
    Wong, Yew-Wah
    ARTIFICIAL ORGANS, 2010, 34 (09) : 788 - 791
  • [5] Experimental evaluation of a novel injection suspended impeller for implantable centrifugal blood pump
    Zhu, Liangfan
    Wu, Yue
    Luo, Yun
    INTERNATIONAL JOURNAL OF APPLIED ELECTROMAGNETICS AND MECHANICS, 2016, 52 (1-2) : 525 - 530
  • [6] Rotordynamic Force Prediction of a Shrouded Centrifugal Pump Impeller-Part I: Numerical Analysis
    Kim, Eunseok
    Palazzolo, Alan
    JOURNAL OF VIBRATION AND ACOUSTICS-TRANSACTIONS OF THE ASME, 2016, 138 (03):
  • [7] Performance discontinuity of a shrouded centrifugal pump impeller: Part 1 - numerical model and flow instability
    Ginter, F
    Staubli, T
    THIRD EUROPEAN CONFERENCE ON TURBOMACHINERY - VOLS A AND B: FLUID DYNAMICS AND THERMODYNAMICS, 1999, 1999 (1A-1B): : 1027 - 1037
  • [8] Rotordynamic Force Prediction of a Shrouded Centrifugal Pump Impeller-Part II: Stability Analysis
    Kim, Eunseok
    Palazzolo, Alan
    JOURNAL OF VIBRATION AND ACOUSTICS-TRANSACTIONS OF THE ASME, 2016, 138 (03):
  • [9] Impeller design for a miniaturized centrifugal blood pump
    Takano, T
    Schulte-Eistrup, S
    Yoshikawa, M
    Nakata, K
    Kawahito, S
    Maeda, T
    Nonaka, K
    Linneweber, J
    Glueck, J
    Fujisawa, A
    Makinouchi, K
    Yokokawa, M
    Nosé, Y
    ARTIFICIAL ORGANS, 2000, 24 (10) : 821 - 825
  • [10] CENTRIFUGAL BLOOD PUMP WITH A MAGNETICALLY SUSPENDED IMPELLER
    AKAMATSU, T
    NAKAZEKI, T
    ITOH, H
    ARTIFICIAL ORGANS, 1992, 16 (03) : 305 - 308