Determination of task group 43 dosimetric parameters for CSM40 137Cs source for use in brachytherapy

被引:2
|
作者
Firoozabadi M.M. [1 ]
Jimabadi E. [1 ]
Ghorbani M. [2 ]
Behmadi M. [3 ]
机构
[1] Department of Physics, Faculty of Sciences, University of Birjand, Birjand
[2] Biomedical Engineering and Medical Physics Department, Faculty of Medicine, Shahid Beheshti University of Medical Sciences, Velenjak, Tehran
[3] Medical Physics Research Center, Mashhad University of Medical Sciences, Mashhad
关键词
Brachytherapy; CSM40 [!sup]137[!/sup]Cs source; Monte Carlo simulation; TG-43 dosimetric parameters;
D O I
10.1007/s12194-017-0440-3
中图分类号
学科分类号
摘要
The CSM40 137Cs source model is currently being used in clinical brachytherapy. According to the recommendations of task group No. 43 (TG-43) of the American Association of Physicists in Medicine, dosimetry parameters of brachytherapy sources should be determined by two independent investigators before their clinical use. The aim of this study was to determine the TG-43 dosimetry parameters for a medium-dose-rate CSM40 137Cs source. The determined dosimetric parameters included the air kerma strength, dose rate constant, radial dose function, and anisotropy function. To determine the source’s dosimetric parameters, the CSM40 source was stimulated by the Monte Carlo N-Particle MCNP code. The TG-43 parameters were compared with the data of Vijande et al. on this source. The results showed that the dosimetry parameters for this source had good agreement with the results of Vijande et al. The dosimetric parameters of the CSM40 source can be used in treatment-planning systems incorporating this source model. The data can also be used for the quality assurance of treatment-planning systems. © 2018, Japanese Society of Radiological Technology and Japan Society of Medical Physics.
引用
收藏
页码:82 / 90
页数:8
相关论文
共 17 条
  • [1] Revision of the dosimetric parameters of the CSM11 LDR Cs-137 source
    Otal, Antonio
    Manuel Martinez-Fernandez, Juan
    Granero, Domingo
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2011, 3 (01) : 36 - 39
  • [2] Revision of the Dosimetric Parameters of the CSM11 LDR Cs-137 Source
    Granero, D.
    Otal, A.
    Fernandez, J. Martinez
    Vijande, J.
    Perez-Calatayud, J.
    Ballester, F.
    MEDICAL PHYSICS, 2011, 38 (06) : 3575 - +
  • [3] A Monte Carlo investigation of the dosimetric characteristics of the CSM11 137Cs source from CIS
    Ballester, F
    Lluch, JL
    Limami, Y
    Serrano, MA
    Casal, E
    Pérez-Calatayud, J
    Lliso, F
    MEDICAL PHYSICS, 2000, 27 (09) : 2182 - 2189
  • [4] AAPM TG-43 formalism for brachytherapy dose calculation of a 137Cs tube source
    Zhang, PP
    Beddar, AS
    Sibata, CH
    MEDICAL PHYSICS, 2004, 31 (04) : 755 - 759
  • [5] Determination of 137Cs dosimetry parameters according to the AAPM TG-43 formalism
    Liu, LZ
    Prasad, SC
    Bassano, DA
    MEDICAL PHYSICS, 2004, 31 (03) : 477 - 483
  • [6] Monte Carlo and experimental derivation of TG43 dosimetric parameters for CSM-type Cs-137 sources
    Pérez-Calatayud, J
    Granero, D
    Casal, E
    Ballester, F
    Puchades, V
    MEDICAL PHYSICS, 2005, 32 (01) : 28 - 36
  • [7] Experimental determination of the Task Group-43 dosimetric parameters of the new I25.S17plus 125I brachytherapy source
    Moutsatsos, Argyris
    Pantelis, Evaggelos
    Papagiannis, Panagiotis
    Batas, Dimos
    BRACHYTHERAPY, 2014, 13 (06) : 618 - 626
  • [8] A comparison of the dosimetric parameters of Cs-137 brachytherapy source in different tissues with water using monte carlo simulation
    Sina, Sedigheh
    Faghihi, Reza
    Meigooni, Ali Soleymani
    Iranian Journal of Medical Physics, 2012, 9 (1 A): : 65 - 74
  • [9] TG-43U1 BASED DOSIMETRIC CHARACTERIZATION OF MODEL 67-6520 CS-137 BRACHYTHERAPY SOURCE
    Ballester, F.
    Meigooni, A.
    Wright, C.
    Koona, R. A.
    Awan, S. B.
    Granero, D.
    Perez-Calatayud, J.
    Otal-Palacin, A.
    RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S222 - S222
  • [10] TG-43U1 based dosimetric characterization of model 67-6520 Cs-137 brachytherapy source
    Wright, C
    Duncan, R
    Rachabatthula, V
    Koona, R
    Meigooni, A
    Baker, C
    MEDICAL PHYSICS, 2005, 32 (06) : 2099 - 2099