Effect of ionic-strength adjusters on the detection of silver ion using ion-selective electrode

被引:1
|
作者
Oh, Seung Yeon [1 ]
Sung, Hwa Kyung [1 ]
Shin, Hyeon Ho [1 ]
Jeong, Uiseok [1 ]
Eom, Ig-chun [2 ]
Kim, Pil-je [2 ]
Kim, Younghun [1 ]
机构
[1] Kwangwoon Univ, Dept Chem Engn, Seoul 139701, South Korea
[2] Natl Inst Environm Res, Risk Assessment Div, Inchon 404708, South Korea
基金
新加坡国家研究基金会;
关键词
Ion Selective Electrode; Silver Ion; Silver Nanoparticles; Toxicity; SPECIATION;
D O I
10.1007/s11814-015-0015-0
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Ion-selective electrodes (ISEs) are used in a wide variety of applications for determining the concentrations of various ions in aqueous solutions. When compared to many other analytical techniques, ISEs are relatively inexpensive, simple to use, and have an extremely wide range of applications and concentrations. In recent reports, cytotoxicity of silver nanoparticles (AgNPs) showed that AgNPs were ionized in the cells and led to causing cell-death. Thus, the measurement of silver ions in aqueous phase is important. In this work, several sodium and potassium salts were considered as ionic-strength adjusters (ISA) and were used in the calibration step to enhance sensitivity of detection the target ion. The result showed that potassium iodine (0.05M KI) acted as a very effective ISA, compared to conventional ISA (5M NaNO3).
引用
收藏
页码:1924 / 1927
页数:4
相关论文
共 50 条