A Standardized Interpolation of Temperature Using Rhodium–Iron Resistance Thermometers Over the Interval 4.2 K to 24.5 K

被引:0
|
作者
W. L. Tew
R. L. Rusby
P. Lin
L. Lipinski
P. P. M. Steur
B. W. A. Ricketson
机构
[1] National Institute of Standards and Technology,Technical Institute of Physics and Chemistry
[2] National Physical Laboratory,undefined
[3] Chinese Academy of Sciences,undefined
[4] Instytut Niskich Temperatur i Badan Strukturalnych,undefined
[5] Istituto Nazionale di Ricerca Metrologica,undefined
[6] Oxford University,undefined
来源
关键词
Interpolation; Low temperature; Rhodium–iron resistance thermometers; Thermometry;
D O I
暂无
中图分类号
学科分类号
摘要
The worldwide history and present state of development of rhodium–iron resistance thermometers (RIRTs) is briefly reviewed. A standardized interpolation method using RIRTs with the nominal composition of 0.5 % Fe (by mole) is presented, with examples using data taken from 60 RIRTs made from a variety of wire batches and sources worldwide over the last 40 years. The parameterization exploits the favorable characteristics of the Cragoe reduced resistance Z(τ)\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$Z(\tau )$$\end{document} and a suitably reduced temperature τ\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$\tau $$\end{document}. A reference function Zref(τ)\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$Z_\mathrm{ref}(\tau )$$\end{document} which approximates the average characteristics of selected wire is derived for use over the interval 0.65 K to 24.5561 K on the ITS-90. The deviations of real RIRT data from this reference function are examined, and simple four-parameter Fourier-series solutions for the resulting deviation curves are presented. Despite the fact that the wire samples may be of different origins or state-of-anneal, it was found that the interpolations are successful for most of the samples studied over the 4.2 K to 24.5561 K interval, at the level of ≈\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$\approx $$\end{document}1 mK\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$\hbox {mK}$$\end{document} standard uncertainty or less. This method would allow for calibrations of most RIRTs over this interval using only six calibration points, permitting an efficiency not achievable using the common least-squares curve-fitting calibration methods. The potential of this formalism for a standardized interpolation scheme using RIRTs is discussed.
引用
收藏
页码:2036 / 2059
页数:23
相关论文
共 50 条
  • [1] A Standardized Interpolation of Temperature Using Rhodium-Iron Resistance Thermometers Over the Interval 4.2 K to 24.5 K
    Tew, W. L.
    Rusby, R. L.
    Lin, P.
    Lipinski, L.
    Steur, P. P. M.
    Ricketson, B. W. A.
    INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2015, 36 (08) : 2036 - 2059
  • [2] Characteristics of Rhodium-Iron Resistance Thermometers and Interpolation Properties from 0.65 K To 24.5561 K
    Rusby, R. L.
    Tew, W. L.
    TEMPERATURE: ITS MEASUREMENT AND CONTROL IN SCIENCE AND INDUSTRY, VOL 8, 2013, 1552 : 409 - 413
  • [3] INTERPOLATION PROCEDURES FOR CERAMIC-ENCAPSULATED RHODIUM IRON ALLOY RESISTANCE THERMOMETERS IN THE TEMPERATURE-RANGE 77-K TO 273-K
    BESLEY, LM
    JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1985, 18 (03): : 201 - 205
  • [4] REPRODUCIBILITY OF GERMANIUM RESISTANCE THERMOMETERS AT 4.2 DEGREES K
    EDLOW, MH
    PLUMB, HH
    JOURNAL OF RESEARCH OF THE NATIONAL BUREAU OF STANDARDS SECTION C-ENGINEERING AND INSTRUMENTATION, 1966, C 70 (04): : 245 - +
  • [5] A SIMPLE ENGINEERING CALIBRATION FOR PLATINUM RESISTANCE THERMOMETERS IN THE TEMPERATURE-RANGE 4.2-K - 273-K
    GERALD, R
    HYMAN, LG
    LADBURY, R
    REZMER, R
    FERNANDEZ, E
    CRYOGENICS, 1983, 23 (02) : 73 - 76
  • [6] Final Report on CCT-K1: Realizations of the ITS-90, 0.65 K to 24.5561 K, using rhodium-iron resistance thermometers
    Rusby, Richard
    Head, David
    Meyer, Christopher
    Tew, Wes
    Tamura, Osamu
    Hill, Kenneth D.
    de Groot, Martin
    Storm, Arnold
    Peruzzi, Andrea
    Fellmuth, Bernd
    Engert, Jost
    Astrov, Dmitri
    Dedikov, Yuri
    Kytin, Gennadi
    METROLOGIA, 2006, 43
  • [7] INTERPOLATION METHODS FOR PLATINUM RESISTANCE THERMOMETERS BETWEEN 13.81-K AND 273.15-K
    KEMP, RC
    KEMP, WRG
    BESLEY, LM
    METROLOGIA, 1981, 17 (02) : 43 - 48
  • [8] INTERPOLATION OF PLATINUM RESISTANCE THERMOMETERS, 20-DEGREES-K TO 273.15-DEGREES-K
    CORRUCCINI, RJ
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1960, 31 (06): : 637 - 640
  • [10] SEMICONDUCTOR RESISTANCE THERMOMETERS FOR THE INTERVAL 300-0.3-DEGREES-K
    ZINOVEVA, KN
    ZARUBIN, LI
    NEMISH, IY
    VOROBKALO, FM
    BOLDAREV, ST
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 1979, 22 (03) : 845 - 847