Numerical comparison of the closing dynamics of a new trileaflet and a bileaflet mechanical aortic heart valve

被引:0
|
作者
Chi-Pei Li
Po-Chien Lu
机构
[1] Tamkang University,Department of Water Resources and Environmental Engineering
来源
关键词
Mechanical heart valve; Valve closing dynamics; Fluid–structure interaction;
D O I
暂无
中图分类号
学科分类号
摘要
The closing velocity of the leaflets of mechanical heart valves is excessively rapid and can cause the cavitation phenomenon. Cavitation bubbles collapse and produce high pressure which then damages red blood cells and platelets. The closure mechanism of the trileaflet valve uses the vortices in the aortic sinus to help close the leaflets, which differs from that of the monoleaflet or bileaflet mechanical heart valves which mainly depends on the reverse flow. We used the commercial software program Fluent to run numerical simulations of the St. Jude Medical bileaflet valve and a new trileaflet mechanical heart valve. The results of these numerical simulations were validated with flow field experiments. The closing velocity of the trileaflet valve was clearly slower than that of the St. Jude Medical bileaflet valve, which would effectively reduce the occurrence of cavitation. The findings of this study are expected to advance the development of the trileaflet valve.
引用
收藏
页码:364 / 374
页数:10
相关论文
共 50 条
  • [1] Numerical comparison of the closing dynamics of a new trileaflet and a bileaflet mechanical aortic heart valve
    Li, Chi-Pei
    Lu, Po-Chien
    JOURNAL OF ARTIFICIAL ORGANS, 2012, 15 (04) : 364 - 374
  • [2] Mechanical Analysis of Mechanical Aortic Heart Valve : Trileaflet versus Bileaflet
    Kiang-ia, Atthasak
    Chatpun, Surapong
    6TH BIOMEDICAL ENGINEERING INTERNATIONAL CONFERENCE (BMEICON 2013), 2013,
  • [3] Direct numerical simulation of the pulsatile flow through an aortic bileaflet mechanical heart valve
    De Tullio, M. D.
    Cristallo, A.
    Balaras, E.
    Verzicco, R.
    JOURNAL OF FLUID MECHANICS, 2009, 622 : 259 - 290
  • [4] Comparison of two spike areas in bileaflet mechanical valve closing sound
    Sugiki, Hiroshi
    Kubota, Suguru
    Matsui, Yoshiro
    Sugiki, Kenji
    JOURNAL OF ARTIFICIAL ORGANS, 2012, 15 (04) : 357 - 363
  • [5] Comparison of two spike areas in bileaflet mechanical valve closing sound
    Hiroshi Sugiki
    Suguru Kubota
    Yoshiro Matsui
    Kenji Sugiki
    Journal of Artificial Organs, 2012, 15 : 357 - 363
  • [6] Computational Hemodynamic Investigation of Bileaflet and Trileaflet Mechanical Heart Valves
    Kuan, Yee Han
    Nguyen, Vinh-Tan
    Kabinejadian, Foad
    Leo, Hwa Liang
    JOURNAL OF HEART VALVE DISEASE, 2015, 24 (03): : 393 - 403
  • [7] Bileaflet mechanical heart valve closing sounds: in vitro classification by phonocardiographic analysis
    Bagno, Andrea
    Anzil, Federico
    Buselli, Roberto
    Pesavento, Elena
    Tarzia, Vincenzo
    Pengo, Vittorio
    Bottio, Tomaso
    Gerosa, Gino
    JOURNAL OF ARTIFICIAL ORGANS, 2009, 12 (03) : 172 - 181
  • [8] A SQUEEZE FLOW PHENOMENON AT THE CLOSING OF A BILEAFLET MECHANICAL HEART-VALVE PROSTHESIS
    BLUESTEIN, D
    EINAV, S
    HWANG, NHC
    JOURNAL OF BIOMECHANICS, 1994, 27 (11) : 1369 - 1378
  • [9] Bileaflet mechanical heart valve closing sounds: in vitro classification by phonocardiographic analysis
    Andrea Bagno
    Federico Anzil
    Roberto Buselli
    Elena Pesavento
    Vincenzo Tarzia
    Vittorio Pengo
    Tomaso Bottio
    Gino Gerosa
    Journal of Artificial Organs, 2009, 12 : 172 - 181
  • [10] A numerical study of hemodynamic effects on the bileaflet mechanical heart valve
    Zbavitel, Jan
    Fialova, Simona
    EXPERIMENTAL FLUID MECHANICS 2018 (EFM18), 2019, 213