Spatial resolution measurement for iterative reconstruction by use of image-averaging techniques in computed tomography

被引:38
|
作者
Urikura A. [1 ,2 ]
Ichikawa K. [3 ]
Hara T. [4 ]
Nishimaru E. [5 ]
Nakaya Y. [1 ]
机构
[1] Department of Diagnostic Radiology, Shizuoka Cancer Center, Sunto, Shizuoka 411-8777, 1007 Shimonagakubo, Nagaizumi
[2] Graduate School of Medical Science, Kanazawa University, Kanazawa, Ishikawa 920-0942
[3] Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, Kanazawa, Ishikawa 920-0942
[4] Department of Medical Technology, Nakatsugawa Municipal General Hospital, Nakatsugawa, Gifu 508-0011
[5] Department of Radiology, Hiroshima University Hospital, Minami-ku, Hiroshima 734-8551
关键词
Computed tomography; Dose reduction; Image noise; Iterative reconstruction; MTF; Spatial resolution;
D O I
10.1007/s12194-014-0273-2
中图分类号
学科分类号
摘要
The purpose of our study was to investigate the validity of a spatial resolution measuring method that uses a combination of a bar-pattern phantom and an image-averaging technique, and to evaluate the spatial resolution property of iterative reconstruction (IR) images with middle-contrast (50 HU) objects. We used computed tomography (CT) images of the bar-pattern phantom reconstructed by the IR technology Adaptive Iterative Dose Reduction 3D (AIDR 3D), which was installed in the multidetector CT system Aquilion ONE (Toshiba Medical Systems, Otawara, Japan). The contrast of the bar-pattern image was set to 50 HU, which is considered to be a middle contrast that requires higher spatial resolution clinically. We employed an image-averaging technique to eliminate the influence of image noise, and we obtained averaged images of the bar-pattern phantom with sufficiently low noise. Modulation transfer functions (MTFs) were measured from the images. The conventional wire method was also used for comparison; in this method, AIDR 3D showed MTF values equivalent to those of filtered back projection. For the middle-contrast condition, the results showed that the MTF of AIDR 3D decreased with the strength of IR processing. Further, the MTF of AIDR 3D decreased with dose reduction. The image-averaging technique used was effective for correct evaluation of the spatial resolution for middle-contrast objects in IR images. The results obtained by our method clarified that the resolution preservation of AIDR 3D was not sufficient for middle-contrast objects. © 2014 Japanese Society of Radiological Technology and Japan Society of Medical Physics.
引用
收藏
页码:358 / 366
页数:8
相关论文
共 50 条
  • [1] Iterative Image Reconstruction Techniques Cardiothoracic Computed Tomography Applications
    Cho, Young Jun
    Schoepf, U. Joseph
    Silverman, Justin R.
    Krazinski, Aleksander W.
    Canstein, Christian
    Deak, Zsuzsanna
    Grimm, Jochen
    Geyer, Lucas L.
    JOURNAL OF THORACIC IMAGING, 2014, 29 (04) : 198 - 208
  • [3] Spatial resolution compensation by adjusting the reconstruction kernels for iterative reconstruction images of computed tomography
    Sugisawa, Koichi
    Ichikawa, Katsuhiro
    Urikura, Atsushi
    Minamishima, Kazuya
    Masuda, Shota
    Hoshino, Takashi
    Nakahara, Akiko
    Yamada, Yoshitake
    Jinzaki, Masahiro
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 74 : 47 - 55
  • [4] Iterative reconstruction techniques in emission computed tomography
    Qi, Jinyi
    Leahy, Richard M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (15): : R541 - R578
  • [5] Image Reconstruction for Ultrasound Computed Tomography by Use of the Regularized Dual Averaging Method
    Matthews, Thomas P.
    Wang, Kun
    Li, Cuiping
    Duric, Neb
    Anastasio, Mark A.
    MEDICAL IMAGING 2017: ULTRASONIC IMAGING AND TOMOGRAPHY, 2017, 10139
  • [6] Spatial resolution of few-view computed tomography using algebraic reconstruction techniques
    Konovalov A.B.
    Kiselev A.N.
    Vlasov V.V.
    Pattern Recognition and Image Analysis, 2006, 16 (2) : 249 - 255
  • [7] Iterative reconstruction techniques for industrial computed tomography: Application and performance
    Arrowood, L.
    Gregor, J.
    Bingham, P.
    MATERIALS EVALUATION, 2007, 65 (11) : 1109 - 1112
  • [8] Bifurcation analysis of iterative image reconstruction method for computed tomography
    Yoshinaga, Tetsuya
    Imakura, Yoshihiro
    Fujimoto, Ken'ichi
    Ueta, Tetsushi
    INTERNATIONAL JOURNAL OF BIFURCATION AND CHAOS, 2008, 18 (04): : 1219 - 1225
  • [9] Iterative Image Reconstruction and Its Role in Cardiothoracic Computed Tomography
    Singh, Sarabjeet
    Khawaja, Ranish Deedar Ali
    Pourjabbar, Sarvenaz
    Padole, Atul
    Lira, Diego
    Kalra, Mannudeep K.
    JOURNAL OF THORACIC IMAGING, 2013, 28 (06) : 355 - 367
  • [10] Proton Computed Tomography: iterative image reconstruction and dose evaluation
    Civinini, C.
    Bonanno, D.
    Brianzi, M.
    Carpinelli, M.
    Cirrone, G. A. P.
    Cuttone, G.
    Lo Presti, D.
    Maccioni, G.
    Pallotta, S.
    Randazzo, N.
    Scaringella, M.
    Romano, F.
    Sipala, V.
    Talamonti, C.
    Vanzi, E.
    Bruzzi, M.
    JOURNAL OF INSTRUMENTATION, 2017, 12