共 50 条
Multi-Pesticide Analysis in Sediment by GC-EI-MS/MS Using Programmed Temperature Vaporization–Large Volume Injection Technique
被引:0
|作者:
Penny Ho-man Lee
Hubert Po-on Tang
机构:
[1] Government Laboratory,Analytical and Advisory Services Division
来源:
关键词:
Tandem mass spectrometry;
PTV-LVI;
Sediment;
Simultaneous determination;
Toxaphene;
Organochlorine pesticides;
D O I:
暂无
中图分类号:
学科分类号:
摘要:
A sensitive, selective and high throughput gas chromatography–tandem mass spectrometry method using programmable temperature vaporization–large volume injection mode (PTV-LVI-GC-MS/MS) for the analysis of 30 organochlorine pesticides (OCPs) including toxaphenes in sediments was developed. The PTV-LV injection settings, viz. inlet temperature, split flow, injection phase time, and injection speed were optimized for 50 µL injection. A significant increase in sensitivity was accomplished as compared with that obtained by the conventional 1 µL cold splitless injection. Average LVI recoveries for OCPs were in the range 58–133 % with low % RSD in instrument precision (<12 %). The method detection limits achieved were 0.04–0.92 µg kg−1. The method recovery ranged from 80 to 120 % with <10 % RSD for more than 83 % of targeted analytes fortified at 10 µg kg−1 in sediments. The PTV-LVI-GC-MS/MS allows simultaneous determination and unambiguous confirmation of trace OCPs and toxaphene congeners, which significantly streamlines and improves the trace organic analysis in the environmental surveillance and monitoring.
引用
收藏
页码:695 / 705
页数:10
相关论文