Pileup Rejection using Estimation Technique for High Resolution Nuclear Pulse Spectroscopy

被引:0
|
作者
Siddavatam, Asma Parveen I. [1 ]
Vaidya, P. P. [1 ]
Nair, J. M. [1 ]
机构
[1] Univ Mumbai, VES Inst Technol, Bombay 400074, Maharashtra, India
关键词
D O I
暂无
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
High resolution spectroscopy systems need accurate measurement of peak height of pulses. The non-zero response time of the detection system leads to overlapping of pulses, thus affecting the overall performance of the system in terms of throughput, system dead time and energy resolution. In this paper, a new pile up detection technique has been described which is based on estimation of peak height of the pulses. For estimating peak value of the pulses; advantage has been taken of the fact that nuclear pulses have constant shape and same peaking time under no pileup condition. The shape of the pulses are exclusively dependent on shaping circuit and only the amplitude of the pulses will change with respect to the energy associated with individual radiation events. Piled up pulses are detected before the peaking time of the pulses. And piled up condition is detected in real time and also it saves lot of computational time. The method developed is easy to implement using basic analog and digital circuits.
引用
收藏
页码:531 / 535
页数:5
相关论文
共 50 条
  • [1] Fast least mean squares method for time determination and pileup rejection in high resolution spectroscopy
    Ripamonti, G
    Geraci, A
    1996 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1-3, 1997, : 474 - 477
  • [2] HIGH-RESOLUTION SPECTROSCOPY USING PICOSECOND PULSE TRAINS
    FERGUSON, AI
    TAYLOR, RA
    PROCEEDINGS OF THE SOCIETY OF PHOTO-OPTICAL INSTRUMENTATION ENGINEERS, 1983, 369 : 366 - 373
  • [3] HIGH-RESOLUTION COHERENCE SPECTROSCOPY USING PULSE TRAINS
    MLYNEK, J
    LANGE, W
    HARDE, H
    BURGGRAF, H
    PHYSICAL REVIEW A, 1981, 24 (02): : 1099 - 1102
  • [4] Pulse pileup correction method for gamma-ray spectroscopy in high radiation fields
    Lee, Minju
    Lee, Daehee
    Ko, Eunbie
    Park, Kyeongjin
    Kim, Junhyuk
    Ko, Kilyoung
    Sharma, Manish
    Cho, Gyuseong
    NUCLEAR ENGINEERING AND TECHNOLOGY, 2020, 52 (05) : 1029 - 1035
  • [7] High-Resolution Terahertz Time-Domain Spectroscopy Using a Wavelet Power Spectrum Estimation Technique
    Kim, Youngehan
    Yee, Dae-Su
    Ye, Jong Chul
    Ahn, Jaewook
    2009 CONFERENCE ON LASERS AND ELECTRO-OPTICS AND QUANTUM ELECTRONICS AND LASER SCIENCE CONFERENCE (CLEO/QELS 2009), VOLS 1-5, 2009, : 764 - +
  • [8] Development of nuclear spectroscopy systems using high resolution semiconductor detector
    Hien, DS
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 373 (03): : 406 - 410
  • [9] Development of nuclear spectroscopy systems using high resolution semiconductor detector
    Hien, Dinh Sy
    Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1996, 373 (03): : 406 - 410
  • [10] EFFECT OF PULSE RATE AND SPECTRUM ON ENERGY RESOLUTION IN NUCLEAR SPECTROSCOPY
    SCHUSTER, HJ
    NUCLEAR INSTRUMENTS & METHODS, 1970, 78 (02): : 274 - +