In-air ion beam analysis with high spatial resolution proton microbeam

被引:2
|
作者
Jaksic, M. [1 ]
Chokheli, D. [1 ]
Fazinic, S. [1 ]
Grilj, V. [1 ]
Skukan, N. [1 ]
Sudic, I. [1 ]
Tadic, T. [1 ]
Anticic, T. [1 ]
机构
[1] Rudjer Boskovic Inst, Bijenicka Cesta 54, Zagreb 10000, Croatia
基金
欧盟地平线“2020”;
关键词
Ion beam analysis; In-air microbeam; IBIC; STIM; Multiple scattering; ANGLE MULTIPLE-SCATTERING; SCREENED COULOMB REGION; EXTERNAL MICROBEAM; GEL-POLYMER; FACILITY; SYSTEM; PIXE; PIGE; IRRADIATION; MICROPROBE;
D O I
10.1016/j.nimb.2015.11.012
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
One of the possible ways to maintain the micrometre spatial resolution while performing ion beam analysis in the air is to increase the energy of ions. In order to explore capabilities and limitations of this approach, we have tested a range of proton beam energies (2-6 MeV) using in-air STIM (Scanning Ion Transmission Microscopy) setup. Measurements of the spatial resolution dependence on proton energy have been compared with SRIM simulation and modelling of proton multiple scattering by different approaches. Results were used to select experimental conditions in which 1 micrometre spatial resolution could be obtained. High resolution in-air microbeam could be applied for IBIC (Ion Beam Induced Charge) tests of large detectors used in nuclear and high energy physics that otherwise cannot be tested in relatively small microbeam vacuum chambers. (C) 2015 Elsevier B.V. All rights reserved.
引用
收藏
页码:185 / 188
页数:4
相关论文
共 50 条
  • [1] Characterisation of the University of Surrey Ion Beam Centre in-air scanning microbeam
    Merchant, MJ
    Mistry, P
    Browton, M
    Clough, AS
    Gauntlett, FE
    Jeynes, C
    Kirkby, KJ
    Grime, GW
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2005, 231 : 26 - 31
  • [2] PROTON MICROBEAM ANALYSIS IN AIR
    COOKSON, JA
    PILLING, FD
    PHYSICS IN MEDICINE AND BIOLOGY, 1976, 21 (06): : 965 - 969
  • [3] Modification of Teflon surface by proton microbeam and nitrogen ion beam
    Kitamura , Akane
    Satoh, Takahiro
    Koka, Masashi
    Kamiya, Tomihiro
    Kobayashi, Tomohiro
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2013, 314 : 82 - 85
  • [4] PROTON AND ION MICROBEAM COLLIMATION FOR IRRADIATION OF BIOLOGICAL SAMPLES IN AIR
    Gaspariunas, M.
    Gervinskas, G.
    Kovalevskij, V.
    Plukiene, R.
    Vaitekonis, S.
    Levenets, V.
    Plukis, A.
    LITHUANIAN JOURNAL OF PHYSICS, 2010, 50 (03): : 363 - 368
  • [5] Development of an in-air high-resolution PIXE system
    Maeda, K
    Hasegawa, K
    Hamanaka, H
    Ogiwara, K
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1998, 134 (3-4): : 418 - 426
  • [6] In-air proton beam irradiation induced radiolysis of methyl orange in aqueous solution
    Huszank, R.
    Nagy, Gy
    Rajta, I.
    Czegledi, A. M.
    RADIATION PHYSICS AND CHEMISTRY, 2021, 180
  • [7] A SIMPLE EXTERNAL-BEAM ION MILLIPROBE SYSTEM FOR IN-AIR PIXE
    MACLAREN, SA
    CORRELL, FD
    HUDDLE, JR
    VANHOY, J
    KULP, WD
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1991, 56-7 (pt 1): : 708 - 711
  • [8] High spatial resolution and high brightness ion beam probe for in-situ elemental and isotopic analysis
    Long, Tao
    Clement, Stephen W. J.
    Bao, Zemin
    Wang, Peizhi
    Tian, Di
    Liu, Dunyi
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2018, 419 : 19 - 25
  • [9] Effect of Range Shifter and Air Gap on In-Air Spot Size for the ProteusPLUS Pencil Beam Scanning (PBS) Proton Beam
    Rana, S.
    Hsi, W.
    Gutierrez, A.
    Coutinho, L.
    Khan, F.
    Fagundes, M.
    Bennouna, J.
    MEDICAL PHYSICS, 2018, 45 (06) : E269 - E269
  • [10] A high resolution magnetic spectrograph for ion beam analysis
    Arnoldbik, WM
    Wolfswinkel, W
    Inia, DK
    Verleun, VCG
    Lobner, S
    Reinders, JA
    Labohm, F
    Boerma, DO
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1996, 118 (1-4): : 566 - 572